Advertisement for orthosearch.org.uk
Results 1 - 20 of 24
Results per page:
Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims. This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported. Methods. This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed. Results. Adjacent to the nail’s telescoping junction, osteolytic changes were observed in bi-planar radiographs of 20/57 segments (35%) after a mean of 9.5 months (95% confidence interval 7.2 to 11.9) after surgery. A total of 8/20 patients with osseous alterations (40%) reported rest and ambulation pain of the lengthened segment during consolidation. So far, 24 Stryde nails were retrieved and in 20 (83%) macroscopic corrosion was observed at the nail’s telescoping junction. Before implant removal 11/20 radiographs (55%) of lengthened segments with these 20 nails revealed osteolysis. Implant retrieval analysis by means of SEM showed pitting and crevice corrosion. EDX detected chromium as the main metallic element of corrosion. Conclusion. Patients are exposed to the risk of implant-related osteolysis of unclear short- and long-term clinical consequences. The authors advocate in favour of an early implant removal after osseous consolidation. Cite this article: Bone Joint Res 2021;10(7):425–436


Bone & Joint Research
Vol. 7, Issue 7 | Pages 476 - 484
1 Jul 2018
Panagiotopoulou VC Davda K Hothi HS Henckel J Cerquiglini A Goodier WD Skinner J Hart A Calder PR

Objectives. The Precice nail is the latest intramedullary lengthening nail with excellent early outcomes. Implant complications have led to modification of the nail design. The aim of this study was to perform a retrieval study of Precice nails following lower-limb lengthening and to assess macroscopical and microscopical changes to the implants and evaluate differences following design modification, with the aim of identifying potential surgical, implant, and patient risk factors. Methods. A total of 15 nails were retrieved from 13 patients following lower-limb lengthening. Macroscopical and microscopical surface damage to the nails were identified. Further analysis included radiology and micro-CT prior to sectioning. The internal mechanism was then analyzed with scanning electron microscopy and energy dispersive x-ray spectroscopy to identify corrosion. Results. Seven male and three female patients underwent 12 femoral lengthenings. Three female patients underwent tibial lengthening. All patients obtained the desired length with no implant failure. Surface degradation was noted on the telescopic part of every nail design, less on the latest implants. Microscopical analysis confirmed fretting and pitting corrosion. Following sectioning, black debris was noted in all implants. The early designs were found to have fractured actuator pins and the pin and bearings showed evidence of corrosive debris. The latest designs showed evidence of biological deposits suggestive of fluid ingress within the nail but no corrosion. Conclusion. This study confirms less internal corrosion following modification, but evidence of titanium debris remains. We recommend no change to current clinical practice. However, potential reuse of the Precice nail, for secondary limb lengthening in the same patient, should be undertaken with caution. Cite this article: V. C. Panagiotopoulou, K. Davda, H. S. Hothi, J. Henckel, A. Cerquiglini, W. D. Goodier, J. Skinner, A. Hart, P. R. Calder. A retrieval analysis of the Precice intramedullary limb lengthening system. Bone Joint Res 2018;7:476–484. DOI: 10.1302/2046-3758.77.BJR-2017-0359.R1


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_6 | Pages 38 - 38
1 May 2021
Iliadis A Timms A Fugazzotto S Edel P Wright J Goodier D Britten S Calder P
Full Access

Introduction. The use of intramedullary lengthening devices is becoming increasingly popular. There are no published data regarding the incidence of venous thromboembolism following intramedullary lengthening and no reports or guidance for current practices on use of thromboprophylaxis. Following a case of post-operative deep vein thrombosis in our institution, we felt that it is important to assess best practice. We conducted this survey to collect data that would describe current practice and help guide consensus for treatment. Materials and Methods. We have identified surgeons across the UK that perform intramedullary lengthening through the British Limb Reconstruction Society membership and a Precise Users database. Surgeons were contacted and asked to respond to an online survey (SurveyMonkey - SVMK Inc.). Responses to thromboprophylaxis regimes employed in their practice and cases of venous thromboembolism were collated. Results. 24 out of 54 surgeons identified responded with a total of 454 cases of intramedullary lengthening (352 femoral and 102 tibial nails) performed over the last 5 years. Only one case of DVT following femoral lengthening was reported. There is wide variability in practice both in terms of thromboprophylaxis risk assessment, choice of medications (20% no pharmacological treatment, 75% Low molecular Weight Heparin, 5% Aspirin) and duration of treatment (0–42 days). The vast majority of surgeons (85%) felt that there was insufficient evidence available to guide their practice. Conclusions. Intramedullary lengthening is a relatively recent and novel surgical treatment. As a result there is limited data available to guide decision making regarding aspects of treatment such as thromboprophylaxis. This is reflected in the wide variation in practice reported in this study. There is both a need and a desire to gather data that will allow us to come to a consensus and to guide safe practice


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_5 | Pages 24 - 24
23 Apr 2024
Thompson E James L Narayan B Peterson N
Full Access

Introduction. Management of deformity involving limb length discrepancy (LLD) using intramedullary devices offers significant benefits to both patients and clinicians over traditional external fixation. Following the withdrawal of the PRECICE nail, the Fitbone became the primary implant available for intramedullary lengthening and deformity correction within our service. This consecutive series illustrates the advantages and complications associated with the use of this device, and describes a novel technique modification for antegrade intramedullary lengthening nails. Materials & Methods. A retrospective cohort review was performed of patient outcomes after treatment with the Fitbone nail at two tertiary referral limb reconstruction services (one adult, one paediatric) between January 2021 to December 2023. Aetiology, indications, initial and final LLD, use of concomitant rail assisted deformity correction (ORDER), removal time and healing index were assessed. Complications of treatment were evaluated and described in detail, alongside technique modifications to reduce the rate of these complications. Results. 21 nails (18 femoral, 2 tibial, 1 humeral) were inserted in 6 adult and 13 paediatric patients. Post-traumatic and congenital/developmental LLD were the most common indications for surgery in the adult and paediatric cohorts respectively. ORDER was employed in 11 cases (9 femurs and 2 tibias). Treatment goals were achieved in all but one case. Complications included superficial infection, locking bolt migration, periprosthetic fracture and component failure. Seven patients required unplanned returns to theatre. Conclusions. The Fitbone nail is an established option for intramedullary limb lengthening, however its use in the UK has been relatively limited compared to the PRECICE until 2021. Our data helps to define its place for limb lengthening and complex deformity correction in both adult and paediatric patients, including in humeral lengthening and retrograde femoral insertion across an open physis. We have identified important potential risks and novel techniques to simplify surgery and avoid complications


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_10 | Pages 19 - 19
1 Jun 2023
Donnan U O'Sullivan M McCombe D Coombs C Donnan L
Full Access

Introduction. The use of vascularised fibula grafts is an accepted method for reconstructing the distal femur following resection of malignant childhood tumours. Limitations relate to the mismatch of the cross-sectional area of the transplanted fibula graft and thel ocal bone, instability of the construct and union difficulties. We present midterm results of a unique staged technique—an immediate defect reconstruction using a double-barrel vascularised fibula graft set in in A-frame configuration and a subsequent intramedullary femoral lengthening. Materials & Methods. We retrospectively included 10 patients (mean age 10 y)with an osteosarcoma of the distal femur, who were treated ac-cording to the above-mentioned surgical technique. All patients were evaluated with regards to consolidation of the transplanted grafts, hypertrophy at the graft-host junctions, leg length discrepancies, lengthening indices, complications as well as functional outcome. Results. The mean defect size after tumour resection was 14.5 cm, the mean length of the harvested fibula graft 22 cm, resulting in a mean (acute) shortening of 4.7 cm (in 8 patients). Consolidation was achieved in all cases, 4 patients required supplementary bone grafting. Hypertrophy at the graft-host junctions was observed in78% of the evaluable junctions. In total 11 intramedullary lengthening procedures in 9 patients had been performed at the last follow up. The mean Muskuloskeletal Society Rating Scale(MSTS) score of the evaluable 9 patients was 85% (57% to 100%)with good or excellent results in 7 patients. Conclusions. A-frame vascularised fibula reconstructions showed encouraging results with respect to defect reconstruction, length as well as function and should therefore be considered a valuable option for reconstruction of the distal femur after osteosarcoma resection. The surgical implementation is demanding though, which is emphasized by the considerable high number of com-plications requiring surgical intervention, even though most were not serious


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_10 | Pages 28 - 28
1 Jun 2023
Musielak B Green N Giles S Madan S Fernandes J
Full Access

Introduction. Intramedullary lengthening devices have been in use in older children with closed /open growth plates with good success. This study aims to present the early experience of the FITBONE nail since withdrawal of the PRECICE nail. Materials & Methods. Retrospective analysis of both antegrade and retrograde techniques were utilized. Only patients where union was achieved and full weight bearing commenced were included. The complication rate, length gained, distraction index, weight bearing index (WBI) as well as mechanical axes were analysed. Results. 14 (7 males, 7 females) of a total of 16 (7 males, 7 females) patients with a mean age of 16.9 years with varied diagnosis of LLD were analysed. The mean length gained was 38 mm with an average distraction index of 0,74 mm/day. WBI in these patients on average was 59,6 days/cm lengthened. 6 complications were observed, including two nonunions (successfully treated) and a knee subluxation. Mechanical axis deviation improved from 13,3 mm to 6 mm on average. Overall there has been a nonsignificant tendency for WBI to decrease (Spearman's rank correlation coefficient −0.47, p=0.08) with increasing number of cases done, while no correlation between length gained and WBI (−0.01, p=0.96, respectively) was observed. Some nuances will be discussed. Conclusions. Limb lengthening with the FITBONE nail is relatively safe and efficient, however no significant change was seen in the outcome with previous motorized nails


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_5 | Pages 9 - 9
23 Apr 2024
Ramlawi AA McClure P Assayag M
Full Access

Introduction. The practice of limb lengthening using intramedullary nails has surged in popularity in recent years. Our study explores the relationship between femur lengthening and overall height gain in adults undergoing cosmetic limb lengthening with telescoping magnetic intramedullary lengthening nails (MILNs). Materials & Methods. Demographic information, pre- and postoperative radiographic data, and secondary outcomes, such as mechanical angles and sagittal alignment, were analyzed for 42 adult femurs MILNs (PRECICE 2, NuVasive, Inc.). Height was assessed with a digital stadiometer. Limb lengthening was defined as the amount of nail distraction seen on a calibrated weight bearing X-ray at consolidation. mLDFA, mMPTA, MAD, AMA, and femoral sagittal bow were evaluated as secondary outcomes. Results. Mean starting height was 163.5 cm (SD 10.4, range 137–179) Post -lengthening mean height increased to 171.1 cm (SD 10.1, range 146.7–185.3). Average height gain was 7.5 cm (SD 1.3 cm, range 5.1–9.5). Average femoral lengthening was 7.3 cm (SD 1.14, range 4.1–10.1). There were no statistically significant difference between height gain and femur lengthening. Secondary outcomes showed reductions in femoral sagittal bow and AMA. Conclusions. We conclude that measuring femur lengthening is a reliable and accurate method of measuring gained height in deformity-free patients undergoing limb lengthening. We further reason that straightening of both femoral and spinal sagittal bow counteracts potential lengthening loss over the anatomical axis, versus mechanical axis lengthening


Introduction. At Sheffield Children's Hospital, treatment of leg length discrepancy is a common procedure. Historically, this has been done with external fixators. With the development in intramedullary technology, internal nails have become the preferred modality for long bone lengthening in the adolescent population. However, it is important to review whether this technology practically reduces the known challenges seen and if it brings any new issues. Therefore, the aim of this review is to retrospectively evaluate the therapeutic challenges of 16 fit-bone intramedullary femoral lengthening's at Sheffield Children's Hospital between 2021–2022. Materials & Methods. The international classification of function (ICF) framework was used to differentiate outcomes. The patient's therapy notes were retrospectively reviewed for themes around structural, activity and participation limitation. The findings were grouped for analysis and the main themes presented. Results. There were 8 males, mean age 17.4 years (range 17–18) and 8 females, mean age 15.9 years (range 14–18). 5 right and 11 left femurs were lengthened. Underlying pathology varied amongst the 16 patients. All patients went into a hinged knee brace post operatively. Structural limitations included: pain, fixed flexion deformity of the knee, loss of knee flexion, quadriceps muscle lag, muscle spasms and gluteal weakness. The primary activity limitation was reduced weight bearing with altered gait pattern. Participation limitations included reduced school attendance and involvement in activities with peers. Access to Physiotherapy from local services varied dramatically. Five of the cohort have completed treatment. Conclusions. Anecdotally, intramedullary femoral lengthening nails have perceived benefits for families compared to external fixators in the adolescent population. However, there remain musculoskeletal and psychosocial outcomes requiring therapeutic management throughout the lengthening process and beyond. Therefore, quantifying these outcomes is essential for measuring the impact on each patient for comparison. To interpret these themes, we need to evaluate the outcomes objectively, this was not done consistently in this review. Future research should look at outcome measures that are sensitive to all aspects of the ICF. With an aim of improving the therapeutic treatment provided and the overall outcome for the children treated


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_5 | Pages 40 - 40
1 Apr 2022
Hafez M Nicolaou N Dixon S Obasohan P Giles S Madan S Fernandes J Offiah A
Full Access

Introduction. Motorised intramedullary lengthening nails are considered more expensive than external fixators for limb lengthening. This research aims to compare the cost of femoral lengthening in children using the PRECICE magnetic lengthening nail with external fixation. Materials and Methods. Patients: Retrospective analysis of 50 children who underwent femoral lengthening. One group included patients who were treated with PRECICE lengthening nails, the other group included patients who had lengthening with external fixation. Each group included 25 patients aged between 11–17 years. The patients in both groups were matched for age. Cost analysis was performed following micro-costing and analysis of the used resources during the different phases of the treatments. Results. : Each group's mean patient age was 14.7 years. Lengthening nails were associated with longer operative times compared to external fixators, both for implantation and removal surgery (P-value 0.007 and <0.0001 respectively). Length of stay following the implantation surgery, frequency of radiographs, frequency of outpatient department appointments were all more favourable with lengthening nails. The overall cost of lengthening nails was £1393 more than external fixators, although this difference was not statistically significant (P-value 0.088). Conclusions. The cost of femoral lengthening with lengthening nails was not significantly higher than the external fixators’ cost. Further research to review the effectiveness of the devices and the quality of life during the lengthening process is crucial for robust health economic evaluation


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_5 | Pages 41 - 41
1 Apr 2022
Hafez M Nicolaou N Offiah A Giles S Madan S Dixon S Fernandes J
Full Access

Introduction. The purpose of this research is to compare the quality of life in children during gradual deformity correction using external fixators with intramedullary lengthening nails. Materials and Methods. Prospective analysis of children during lower limb lengthening. Group A included children who had external fixation, patients in group B had lengthening nails. Patients in each group were followed up during their limb reconstruction. CHU-9D and EQ-5DY instruments were used to measure quality of life at fixed intervals. The first assessment was during the distraction phase (1 month postop.), the second was during the early consolidation phase (3 months postop.) and the final one was late consolidation phase (6–9 months depends on the frame time). Results. Group B patients reported significantly better utility compared to Group A. This was observed during all the stages of the treatment. Group B children were less worried (P 0.004), less sad (P 0.0001), less pain (p <0.0001), less tired (P 0.0002), better school work (P0.0041), better sleep (p 0.016), more able to do sports activities (p 0.004) and, they were more independent (p <0.0001) compared to group B. QALYS was better for the nails group compared to external fixation group 0.44 compared to 0.36 for external fixators. Conclusions. Lengthening nails had the potential to improve the quality of life and utility compared to external fixation. This will help further economic evaluation to measure ICER to further explore the cost effectiveness of these devices


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_8 | Pages 4 - 4
1 May 2018
Timms A Roskrow T Doyle H Edel P Fugazzotto S Goodier W Calder P
Full Access

Introduction. Muscle stiffness and joint contractures are common complications of limb lengthening. Authors have demonstrated less permanent soft tissue complications with intramedullary lengthening than external fixation. Our aim was to evaluate the joint response following intramedullary femoral lengthening and need for physiotherapy and alteration to rate/rhythm of lengthening. Method. A retrospective review of documentation for all femoral Precice nails in our centre inserted between 2012 and 2017. This involved 98 nails (68 antegrade, 30 retrograde) in 88 patients (59 males, 29 females) with a mean age of 32 years (range 12–69 yrs). We excluded cases where there was no documentation regarding Range of Movement (ROM). Bilateral lengthenings were recorded as separate cases. This left 50 antegrade, 16 retrograde cases with hip ROM data and 55 antegrade, 26 retrograde cases with Knee ROM data. Results. Hip. In the antegrade group 20 cases (39%) developed no stiffness throughout treatment. ROM decreased in 61% (n=30) of cases and the median length at which this occurred was 3cm. Where a retrograde nail was inserted 62.5% of cases (n=10) maintained normal range of motion. The median length at which ROM decreased was 3.25cm, this occurred in 6 cases (37.5%). . Knee. Where the antegrade nail was inserted, 22 cases (40%) did not develop stiffness. There was reduced ROM in 33 cases (60%), occurring at a median distance of 2cm. The median distance at which reduced ROM occurred with the retrograde nails was 1.5cm, this occurred in 23 cases (88.5%). 3 cases did not develop stiffness. All cases regained full ROM. Conclusion. Although there was no permanent loss, Joint Stiffness still occurs with intramedullary limb lengthening and there remains the need for regular physiotherapy. There are no data from other institutions for comparison. Rehabilitation guidelines and a proforma to accurately monitor patients ROM throughout treatment have been developed


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_6 | Pages 25 - 25
1 May 2021
Stoddart M Elsheikh A Wright J Goodier D Calder P
Full Access

Introduction. Pixel Value Ratio (PVR) is a radiographic measure of the relative density of the regenerate to the adjacent bone. This has been reported as an objective criterion for regenerate healing and a guide for when to allow full weight bearing (FWB) in lengthening with intramedullary telescopic nails. The threshold for which magnitude of PVR is adequate to allow bearing full weight is not yet agreed. The aim of this study was to identify from our cohort of adult limb lengthening patients the time to FWB following lengthening, the PVR at this point, and how this compared with the recommended values in the literature. Materials and Methods. A retrospective database review identified 30 adult patients treated with the PRECICE femoral nail by two senior authors. Time from completion of lengthening to instruction to fully weight bear was noted. The PVR was calculated for each cortex on plain radiographs taken at each postoperative visit following completion of lengthening. Significance was set at p <0.05. Results. The median age was 30.5 years (IQR 22.5 – 42.5), 19 male and 11 female patients were included. The underlying cause of shortening was post traumatic in 12 (40%) with the remaining due to a wide variety of causes including congenital, infective, and idiopathic leg length discrepancies. Twenty-two nails were antegrade and the mean distance lengthened was 42 mm. Median time from completion of lengthening to full weight bearing was 42 days (IQR 28 – 69). The overall mean PVR at FWB was 0.79. Each surgeon had differing protocols for weightbearing, however there was no statistical difference between time to weight bearing, nor the mean PVR at FWB. There were no implant failures, shortening, or regenerate fractures. Conclusions. We report on our cohort of adult patients treated with intramedullary lengthening nails with a mean PVR of 0.79 at time of full weightbearing. This is considerably lower than the values reported in the literature ranging from 0.90 to 0.93. We therefore suggest that weightbearing following lengthening nails can be permitted earlier than previously reported without an increased risk of complication


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_5 | Pages 22 - 22
23 Apr 2024
Laufer A Frommer A Gosheger G Toporowski G Rölfing JD Antfang C Roedl R Vogt B
Full Access

Introduction. Coronal malalignment and leg length discrepancies (LLD) are frequently associated. Temporary hemiepiphysiodesis (tHED) is commonly employed for the correction of limb malalignment in skeletally immature patients. For treatment of LLD greater than 2 cm, lengthening with intramedullary legnthening nails is a safe and reliable technique. However, the combined application of these approaches in skeletally immature patients has not yet been investigated. Materials & Methods. Retrospective radiological and clinical analysis of 25 patients (14 females, 11 males) who underwent intramedullary femoral lengthening with an antegrade PRECICE® lengthening nail as well as tHED of the distal femur and / or proximal tibia between 2014 and 2019. tHED was conducted by implantation of flexible staples (FlexTack™) either prior (n = 11), simultaneously (n = 10), or subsequently (n = 4) to femoral lengthening. The mean follow-up period was 3.7 years (±1.4). Results. The median initial LLD was 39.0 mm (35.0–45.0). 21 patients (84%) presented valgus and 4 (16%) showed varus malalignment. Leg length equalization was achieved in 13 patients at skeletal maturity (62%). The median LLD of patients with a residual LLD > 10 mm was 15.5 mm (12.8–21.8). Limb realignment was obtained in nine of seventeen skeletally mature patients (53%) in the valgus group, and in one of four patients (25%) in the varus group. Conclusions. The combination of antegrade femoral lengthening and tHED can efficiently correct LLD and coronal limb malalignment in skeletally immature patients. Nevertheless, achieving limb length equalization and realignment may render difficult in cases of severe LLD and angular deformity. Furthermore, the reported techniques ought to be thoroughly planned and executed and require regular clinical and radiological examinations until skeletal maturity to avoid - or timely detect and manage - adverse events such as overcorrection and rebound of deformity


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 402 - 402
1 Jul 2010
Monga P Wilkes RA
Full Access

Introduction: Limb lengthening using external fixation may be associated with problems such as pin-track infections, poor patient acceptance, muscle transfixation, secondary axial deformity and re-fractures. Intramedullary lengthening nails have been designed to address these issues. Aim: To review the outcomes for femoral limb lengthening in adults managed by intramedullary lengthening nails. Materials and Methods: A retrospective review was undertaken for 8 femoral lengthening procedures performed using intra-medullary lengthening nails over a three-year period. The average age of our patients was 34 years and the average duration of follow up was 26.5 months (Range 8 to 40 months). Either an Albizzia nail (5 femurs) or an ISKD (3 femurs) nail was used for the procedure. Results: Target lengthening was achieved in 7 out of 8 femurs with a average of 38 mm (Range 19 to 70 mm) length gained. The distraction index (length gained per day) was 0.68 on an average and the consolidation index (length of bone consolidating per day) being 0.27 on an average. Premature consolidation in 4 cases, runaway acute lengthening in one patient, prominent metalwork in 4 patients and a bent nail were frequent obstacles and meant multiple visits to theatre. Conclusions: The desired femoral lengthening is achievable using intra-medullary lengthening nails, thereby avoiding problems associated with callostasis using external fixation methods. It is however, important to counsel patients regarding possibilities of significant obstacles and multiple visits to theatre during the process


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_II | Pages 140 - 140
1 Feb 2004
Escribá-Urios I Fidalgo A Embodas M Crusi J
Full Access

Introduction and Objectives: Our aim is to analyze results using an Albizzia® gradual lengthening intramedullary nail in the femur. Materials and Methods: From October 1997 to November 2000, 7 femoral lengthenings were performed on 5 patients (2 were bilateral) in our unit, using an intramedullary lengthening nail. Average age was 19 years (15–22). Aetiologies included congenital deformity with symmetric shortening (1 case), fibrous cortical defects (1 case), iatrogenic shortening secondary to trauma (1 case), and 2 cases of symmetric dwarfism (idiopathic drawfism, Turner’s syndrome). Distraction was achieved at 15 cycles/day (1 mm/day). Clinical and radiographic results were evaluated using the Paley criteria, considering the indices of consolidation and distraction. Average follow-up time was 48 months (30–60 months). Results: Average lengthening obtained was 5.7 cm (4–7 cm), with a distraction period of 99 days (45–214 days) and a distraction index of 0.71 mm/day (17 days/1 cm). In bilateral cases, the distraction index was 0.57 mm/day. Consolidation time was 187 days with a consolidation index of 33 days/cm. Complications included 3 mechanical (2 broken screws and one intramedullary saw failure); one involving bone (1 intraoperative fracture), and one case requiring spinal anaesthesia to achieve lengthening in the first few days. No nerve, joint, or infection-related complications were observed. Average duration of surgery was 2 hours 40 minutes. Discussion and Conclusions: The use of a gradual intramedullary lengthening nail yields good results, since it is a more stable system which minimises complications compared to external fixators. It is also a more comfortable procedure for the patient


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_6 | Pages 29 - 29
1 May 2021
Rouse B Giles S Fernandes J
Full Access

Introduction. We have previously published limb lengthening using external fixation in pathological bone diseases. We would like to report a case series of femoral lengthening using the PRECICE system in a similar pathological group especially looking at it's feasibility and complications. Materials and Methods. This is a case series of four patients, two patients with osteogenesis imperfecta and two with Ollier's disease, who underwent femoral lengthening via distraction osteogenesis using the PRECICE intramedullary nail system. It was a retrospective study from a prospective database from clinical records and radiographs. Results. The mean age at the time of surgery was 15.5 years, the mean preoperative leg length discrepancy was 30mm, and the mean distraction distance achieved was 28.75mm. Since these patients were of shorter heigh, limb lengthening was considered. All 4 patients had successful insertion of the nail. The outcomes noted from the 4 patients are collated, with several complications occurring including delayed femoral union, fixed flexion deformity of the hip, persisting pain and quadriceps weakness. Those with Ollier's disease underwent an increased rate of distraction to prevent premature healing. The implications of long-term bisphosphonate therapy in OI are discussed with regards to the risk of delayed femoral union and intra-operative fracture. Conclusions. Intramedullary femoral lengthening in pathological bone disease is possible, but the surgeon needs to give attention to certain details. The regenerate formation is based on the background pathology irrespective of the hardware used. There is much more compliance with the nail technique


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_13 | Pages 1 - 1
1 Jun 2017
Panagiotopoulou V Davda K Hothi H Henckel J Cerquiglini A Goodier W Skinner J Hart A Calder P
Full Access

Introduction. The Precice nail is the latest intramedullary lengthening nail with excellent early outcomes. Implant complications have led to modification of the nail design. The aim of this study was to perform a retrieval study of Precice nails following lower limb lengthening. To assess macroscopic and microscopic changes to the implants and assess differences following design modification, with identification of potential surgical, implant and patient risk factors. Method. 15 nails were retrieved from 13 patients following lower limb lengthening. Macroscopic and microscopic surface damage to the nails were identified. Further analysis included radiology and micro-CT prior to sectioning. The internal mechanism was then analysed with Scanning Electron Microscopy and Energy Dispersive X-ray Spectroscopy to identify corrosion. Results. 7 male and 3 females underwent 12 femoral lengthenings, 9 antegrade and 3 retrograde. 3 females underwent tibial lengthening. All patients obtained the desired length with no implant failure and full regenerate consolidation. Surface degradation was noted on the telescopic part of every nail design, less on the latest implants. Microscopic analysis confirmed fretting and pitting corrosion. Following sectioning black debris was noted in all implants. The early designs were found to have fractured actuator pins and the pin and bearings had evidence of corrosive debris. The latest designs had evidence of biological deposits suggestive of fluid ingress within the nail. Conclusion. This study suggests fluid ingress occurs with every generation of Precice nail despite modifications. The presence of biological fluid could be an early warning sign of potential corrosion. This in theory could lead to actuator pin fracture and implant failure. The clinical relevance is the potential re-use of a “dormant” nail in patients requiring secondary limb segment lengthening. Retraction of the nail in-situ and re-use for further lengthening requires careful consent for potential implant failure


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_II | Pages 94 - 94
1 Feb 2012
Acharya A Guichet J Hobson P
Full Access

To assess the effect on knee motion of gradual femoral lengthening using an intramedullary nail, between 1994 and 2003, 27 non-achondroplastic patients had bilateral femoral lengthening using the Albizzia nail. Vigorous post-operative physiotherapy was the norm. Knee motion recorded at various stages pre and post-operatively was compared. For an average gain of 6 cm the mean flexion during lengthening was 119. By final follow-up all patients had regained pre-operative range of motion. No tenotomies or joint manipulations were required. Good knee motion can be maintained during femoral lengthening using an intramedullary lengthening device


Bone & Joint 360
Vol. 1, Issue 3 | Pages 28 - 30
1 Jun 2012

The June 2012 Children’s orthopaedics Roundup. 360. looks at; open reduction for DDH; growing rod instrumentation for scoliosis; acute patellar dislocation; management of the relapsed clubfoot; clubfoot in Iran; laughing gas and fracture manipulation; vascularised periosteal fibular grafting for nonunion; slipped upper femoral epiphysis; intramedullary leg lengthening and orthopaedic imaging and defensive medicine


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_23 | Pages 11 - 11
1 May 2013
Calder P McGrath A Chasseaud M Timms A Goodier W
Full Access

We present the results of a new non-invasive lengthening nail enabling accurate control of the lengthening process and joint rehabilitation. Introduction. The use of intramedullary lengthening nails have gained popularity as they reduce common complications associated with external fixators, including infection, joint stiffness, bone regenerate deformity, late fracture and patient implant acceptance. Current nails however are associated with complications including implant breakage, mechanical failure, runaway nail and requiring MUA to restart or obtain segment lengthening. The Precice nail incorporates magnet technology with a hand held device allowing non-invasive lengthening. The nail is also reversible allowing shortening if required. Physiotherapy can continue throughout treatment to maintain joint range of motion without concern of uncontrolled nail runaway. The lengthening is axial reducing shear/torsional forces on the regenerate. Method. The lengthening of 4 femora was undertaken in 3 patients, mean age 34 yrs for post-traumatic shortening and short stature. A standard technique included an Ilizarov corticotomy followed by a 6 day latent period. Patients were mobilised partial weight bearing and knee range of motion maintained. The femora were lengthened one third of a millimetre three times per day. Radiographic and clinical review was performed every 2 weeks. Results. The desired femoral length was obtained in all patients, mean 5.6 cms (4.5–6.5). There were no complications of infection, poor bone regenerate or premature consolidation. One patient undergoing bilateral femoral lengthening underwent surgical release of a tight Tensor Fascia Lata following completion of lengthening due to hip flexion contracture. One patient required slowing of the lengthening rate due to knee flexion contracture. Physiotherapy corrected the deformity and the normal lengthening rate could be resumed. Conclusion. The Precice nail is a new device that offers accurate control of the lengthening process and rehabilitation. This should reduce complications of poor regenerate formation and soft tissue contractures