Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 46 - 46
1 Sep 2012
Fong J Dunbar MJ Wilson DA Hennigar A Francis P Glazebrook M
Full Access

Purpose

The purpose of this study was to assess the biomechanical stability of the a total ankle arthroplasty system using longitudinal migration (LM) and inducible displacement (ID) measures. This study is the first study of its kind to assess total ankle arthroplasty (TAA) implant micromotion using model-based radiostereometric analysis (MBRSA).

Method

Twenty patients underwent TAA that implanted the Mobility(TM) (DePuy, Warsaw IN). The mean (SD) age was 60.4 (12.5) and BMI was 29.1 (2.8) kg/m2. One surgeon performed all surgeries. All patients included in this study had given informed consent. Capital Health Research Ethics Board had approved this study.

Uniplanar medial-lateral RSA X-ray exams were taken postop (double exam), at six wk, three mth, six mth, one yr and two yr followup times using a supine, unloaded position. Standing medial-lateral exams were taken at three mth, six mth, one yr and two yr followup intervals.

LM and ID micromotions were assessed using Model-based RSA 3.2 software (Medis specials, Leiden, The Netherlands). Implant micromotions (x, y, z, Rx, Ry, Rz, MTPM) were determined and assessed for each subject using model-based pose estimation, and the implant-based coordinate system. The Elementary Geometric Shapes module from the Model-based RSA 3.2 software was used to assess the micromotion of the tibial component spherical tip due to implant symmetry.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 80 - 80
19 Aug 2024
Malchau H
Full Access

The painful hip without obvious clinical or radiographic signs of complications is a well-known scenario for surgeons. The clinical tools we have access to currently lack a dynamic test for detecting early signs of motion between implant and bone. A new software, Sectra IMA, has a potential to facilitate diagnosis of early implant loosening by analysis of paired CT exams. In clinical practise the two scans are acquired by endpoint of a possible motion, “a provocation CT”, for example maximal external and internal rotation in a CT hip examination. 20 years of research by Olivecrona and Weidenhielm is the scientific background for the technique. Early results are presented by Sandberg et al 2022. To further validate and create clinical evidence more extensive clinical studies is needed. A proposal for a multicentric, global clinical study will be presented for further discussion. A new technique for diagnosis of acetabular cup loosening using computed tomography: preliminary experience in 10 patients. Henrik Olivecrona, Lotta Olivecrona, Lars Weidenhielm, Marilyn E Noz, Jocelyn Kardos, Gerald Q Maguire Jr, Michael P Zeleznik, Peter Aspelin. Acta Orthop. 2008 Jun;79(3):346-53. doi: 10.1080/17453670710015247. Inducible displacement CT increases the diagnostic accuracy of aseptic loosening in primary total hip arthroplasty. Sandberg O, Carlsson S, Harbom E, Cappelen V, Tholén S, Olivecrona H, Wretenberg P. Acta Orthop. 2022 Oct 31;93:831-836. doi: 10.2340/17453674.2022.5240.PMID: 36314542


Bone & Joint Open
Vol. 5, Issue 1 | Pages 20 - 27
17 Jan 2024
Turgeon TR Vasarhelyi E Howard J Teeter M Righolt CH Gascoyne T Bohm E

Aims. A novel enhanced cement fixation (EF) tibial implant with deeper cement pockets and a more roughened bonding surface was released to market for an existing total knee arthroplasty (TKA) system.This randomized controlled trial assessed fixation of the both the EF (ATTUNE S+) and standard (Std; ATTUNE S) using radiostereometric analysis. Methods. Overall, 50 subjects were randomized (21 EF-TKA and 23 Std-TKA in the final analysis), and had follow-up visits at six weeks, and six, 12, and 24 months to assess migration of the tibial component. Low viscosity bone cement with tobramycin was used in a standardized fashion for all subjects. Patient-reported outcome measure data was captured at preoperative and all postoperative visits. Results. The patient cohort mean age was 66 years (SD seven years), 59% were female, and the mean BMI was 32 kg/m. 2. (SD 6 kg/m. 2. ). Mean two-year subsidence of the EF-TKA was 0.056 mm (95% confidence interval (CI) 0.025 to 0.086) versus 0.006 mm (95% CI -0.029 to 0.040) for the Std-TKA, and the two-year maximum total point motion (MTPM) was 0.285 mm (95% upper confidence limit (UCL) ≤ 0.363) versus 0.346 mm (95% UCL ≤ 0.432), respectively, for a mean difference of -0.061 mm (95% CI -0.196 to 0.074). Inducible displacement also did not differ between groups. The MTPMs between 12 and 24 months for each group was below the published threshold of 0.2 mm for predicting early aseptic loosening (p < 0.001 and p = 0.001, respectively). Conclusion. Both the enhanced fixation and the standard tibial implant design showed fixation with a predicted low risk of long-term aseptic loosening. Cite this article: Bone Jt Open 2024;5(1):20–27


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 26 - 26
1 Oct 2018
Dunbar MJ Laende E Richardson CG
Full Access

Introduction. Cementless fixation in total knee arthroplasty has been proposed to offer advantages long-term once osteointegration has occurred as there is no substrate between the implant and the bone to fail. Radiostereometric analysis (RSA) is a useful tool to study fixation, but typically focused on early migration in the first two post-operative years. Few studies have looked at 10-year RSA migration in cementless fixation and those that have contain small numbers of subjects. The objective of this study was to compare implant migration and inducible displacement between cemented and cementless TKA at 10 years and to compare the 10-year migration to the 2-year data in an effort to validate the predictive modelling of RSA. Methods. Subjects who had previously participated in RSA migration studies with 2-year follow-up were recruited to return for a long-term follow-up exam, at least 10 years from their surgery. The implants under study included two cemented designs from two manufacturers and one porous metal monoblock cementless design. At the 10-year visit, subjects had supine RSA exams to determine long-term migration as well as a loaded exam (single leg stance) to determine inducible displacement. Differences between cemented and cementless groups were evaluated with the Mann Whitney U test and Spearman's rank correlation coefficients were calculated for early and late migrations. Significance was set at p < 0.05. Results. Seventy-five subjects were available for long-term follow-up, with average time since surgery of 12 years. This cohort contained 51 women and 24 men with cemented tibial components in 53 cases (37 female) and cementless tibial components in 22 cases (14 female). At the time of surgery, the subjects were 62±7 years old with BMIs of 33±6 m/kg2 (mean±standard deviation). Median migration at the long-term follow-up was 0.6 mm (MTPM; range 0.2–2.8 mm) and was not different between the cemented and cementless groups (p = 0.9, Mann Whitney U Test). Inducible displacement at 10 years was significantly lower for the cementless implants (p<0.001, Figure 1). Migration at one, two, and 10 years did not correlate with inducible displacement at 10 years. However, migration at one year and two years did correlate with long-term migration, with the strongest correlation at two years (Spearman's rank correlation coefficient for all components = 0.74, p < 0.001, Figure 2). Conclusion. Although long-term migration was not different for cemented or cementless (porous metal monoblock) tibial components, inducible displacement at the 10-year visit was significantly lower for these cementless components, suggesting superior fixation. Additionally, long-term migration was strongly correlated to two-year migration, regardless of fixation. These findings support the predictive value of short-term migration in determining long-term fixation. For any figures or tables, please contact authors directly


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 447 - 447
1 Sep 2009
Keeling P Prendergast P Lennon A O’Reilly P Britton J Kenny P
Full Access

The cement-in-cement femoral revision is a possible method of reducing complications. During recent research on this revision it was observed that a number of the inner cement contained macropores. It was hypothesized that porosity of the mantle influenced the subsidence and inducible displacement of the revision stems. The aim was to calculate the porosity and assess its relationship to the above factors. Primary cement mantles were formed by cementing a stem into sections of tubular steel. At this stage, the specimen was chosen to be in a test or a control group. If in the test group, it underwent a fatigue of 1 million cycles. This was carried out in a fatigue machine mounted with a specifically designed rig. If in the control group, no such fatigue was undertaken. Into these fatigued and unfatigued mantles, the cement-in-cement procedure was performed. Both groups underwent a fatigue of again 1 million cycles. Subsidence and inducible displacement was recorded. The composites were then sectioned and photographed. The images underwent image analysis to calculate the porosity. Multiple regression and a general linear model showed subsidence was inversely correlated to the porosity of the “fresh cement” in Gruen zones 3 and 5 (p = 0.021, R2 = 0.36). This relationship was not expected. The reason could be related to the fact that the migration of the stems in each separate direction was not monitored. Inducible displacement was inversely correlated to porosity of the inner cement, again in Gruen zones 3 and 5 (p = 0.001, R2 = 0.61). A possible explanation is that the stem was able to subside more due to the higher porosity and find a more stable position. The subsidence and inducible displacement of these stems is influenced by porosity, specifically by the porosity of the distal inner cement


Bone & Joint Open
Vol. 2, Issue 1 | Pages 48 - 57
19 Jan 2021
Asokan A Plastow R Kayani B Radhakrishnan GT Magan AA Haddad FS

Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes.

Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty.

This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice.

Cite this article: Bone Jt Open 2021;2(1):48–57.