Bone & Joint Research
Vol. 7, Issue 8 | Pages 501 - 507
1 Aug 2018
Phan C Nguyen D Lee KM Koo S

Objectives. The objective of this study was to quantify the relative movement between the articular surfaces in the tibiotalar and subtalar joints during normal walking in asymptomatic individuals. Methods. 3D movement data of the ankle joint complex were acquired from 18 subjects using a biplanar fluoroscopic system and 3D-to-2D registration of bone models obtained from CT images. Surface relative velocity vectors (SRVVs) of the articular surfaces of the tibiotalar and subtalar joints were calculated. The relative movement of the articulating surfaces was quantified as the mean relative speed (RS) and synchronization index (SI. ENT. ) of the SRVVs. Results. SI. ENT. and mean RS data showed that the tibiotalar joint exhibited translational movement throughout the stance, with a mean SI. ENT. of 0.54 (. sd. 0.21). The mean RS of the tibiotalar joint during the 0% to 20% post heel-strike phase was 36.0 mm/s (. sd. 14.2), which was higher than for the rest of the stance period. The subtalar joint had a mean SI. ENT. value of 0.43 (. sd. 0.21) during the stance phase and exhibited a greater degree of rotational movement than the tibiotalar joint. The mean relative speeds of the subtalar joint in early (0% to 10%) and late (80% to 90%) stance were 23.9 mm/s (. sd. 11.3) and 25.1 mm/s (. sd 9.5). , respectively, which were significantly higher than the mean RS during mid-stance (10% to 80%). Conclusion. The tibiotalar and subtalar joints exhibited significant translational and rotational movement in the initial stance, whereas only the subtalar joint exhibited significant rotational movement during the late stance. The relative movement on the articular surfaces provided deeper insight into the interactions between articular surfaces, which are unobtainable using the joint coordinate system. Cite this article: C-B. Phan, D-P. Nguyen, K. M. Lee, S. Koo. Relative movement on the articular surfaces of the tibiotalar and subtalar joints during walking. Bone Joint Res 2018;7:501–507. DOI: 10.1302/2046-3758.78.BJR-2018-0014.R1


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 50 - 50
1 Apr 2019
Dharia M Wentz D Mimnaugh K
Full Access

INTRODUCTION. Tibiofemoral contact at the base of the articular surface spine in posterior-stabilized total knee arthroplasty (TKA) implants can lead to spine fracture [1]. Revision TKA implants also have an articular surface spine to provide sufficient constraint when soft tissues are compromised. While some revision TKA designs have metal reinforcement in the articular surface spine, others rely solely on a polyethylene spine. This study used finite element analysis (FEA) to study the effect of metal reinforcement on stresses in the spine when subjected to posteriorly directed loading. METHODS. Two clinically successful Zimmer Biomet revision TKA designs were selected; NexGen LCCK with metal reinforcement and all-poly Vanguard SSK. The largest sizes were selected. FEA models consisted of the polyethylene articular surface and a CoCr femoral component; LCCK also included a CoCr metal reinforcement in the spine. A 7° and 0° tibial slope, as well as 3° and 0.7° femoral hyperextension, were used for the LCCK and SSK, respectively. A posteriorly directed load was applied to the spine through the femoral component (Figure 1). The base of the articular surface was constrained. The articular surfaces for both designs are made from different polyethylene materials. However, for the purpose of this study, to isolate the effect of material differences on stresses, both were modeled using conventional GUR1050 nonlinear polyethylene material properties. Femoral component and metal reinforcement were modeled using linear elastic CoCr properties. Additionally, the LCCK was reanalyzed by replacing the metal reinforcement component with polyethylene material, in order to isolate the effect of metal reinforcement for an otherwise equivalent design. Frictional sliding contact was modeled between the spine and femoral/metal reinforcement components. Nonlinear static analyses were performed using Ansys version 17 software and peak von mises stresses in the spine were compared. RESULTS. Peak von mises stresses were predicted towards the base of the anterior aspect of the spine in both designs (Figure 2). In LCCK, the high stresses were also predicted on the medial and lateral edges of the anterior spine, matching the tibiofemoral contact (Figure 3). The LCCK with metal reinforcement design predicted 14% and 31% lower stress than LCCK and SSK all-poly designs. DISCUSSION. Clinical reports of spine fracture in TKA highlight the need for further understanding of the biomechanics of spine loading. Here, through comparison of two clinically successful devices, the effect of multiple design factors was quantified. Inclusion of metal reinforcement in the spine, as well as differences in the conforming geometry between the femoral component and the articular surface, resulted in a 31% decrease in polyethylene stress for the LCCK as compared to the all-poly SSK; of which only 16% was attributed to the metal reinforcement. Further improvements to articular surface design, as well as polyethylene material advances, have the potential to result in all-poly designs with strength characteristics equivalent to or exceeding those of designs with metal reinforcement


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 46 - 46
1 Mar 2017
Nagamine R Weijia C Todo M Osano K Takayama M Kawasaki M Kaneyama R
Full Access

Introduction. The hip-knee-ankle (HKA) angle between the mechanical axis of the femur (FM) and the mechanical axis of the tibia (TM) is the standard parameter to assess the coronal alignment of the lower extremity. TM is the line between the center of the tibial spines notch (Point T) and the center of the tibial plafond. However, this theory is based on the premise that TM coincides the anatomical axis of the tibia (TA). Fig.1a shows typical varus knee with medial shift of the tibial articular surface. In this case, TM does not coincide TA. Fig. 2 demonstrates the error of HKA angle when Point T locates medial to TA. Fig.2a shows normal alignment. Fig.2b shows varus alignment. Fig. 2c shows the tibia with medial shift of the tibial articular surface. The tibia has 7 degrees varus articular inclination in Fig.2b and 2c. However, HKA angle is 0 degree in Fig.2c. HKA angle underestimates varus deformity in knees with medial shift of the tibial articular surface. However, the degree of medial shift of the tibial articular surface is obscure. In this study, detailed anatomical configuration of the proximal tibia was evaluated. The effect of the value of HKA angle on the coronal alignment in TKA was then discussed. Methods. This study consists of 117 knees. On the AP view radiograph of the tibia, three distance and two angle parameters were measured. Those were tibial articular surface width, distance between medial edge of the tibial articular surface and Point T, distance from TA to Point T. Angle between TM and TA, and the varus inclination angle of the tibial articular surface relative to the perpendicular line to TA. Results. The mean width of the tibial articular surface was 87.8mm. The mean distance between Point T and medial edge of the articular surface was 43.6mm. Point T located at the center of the tibial articular surface. The mean distance from TA to Point T was 5.6mm. The mean angle between TM and TA was 1.0 degrees. The inclination angle of the tibial articular surface was 8.2 degrees. Discussion. The results clearly showed that varus knees had medial shift of the tibial articular surface. In such knees, the ankle shifts laterally relative to the Point T and HKA angle underestimates the varus deformity. The value of HKA angle also influences the evaluation of the coronal alignment of the lower extremity in TKA (Fig. 3). When the tibial tray is set based on the tibial articular surface width in order to cover the cut surface of the tibia, HKA angle shows the alignment as valgus when the tibial tray is set perpendicular to TA (Fig. 3a). In order to obtain zero degree of HKA angle, the tibial tray should be set perpendicular to TM. This alignment is varus (Fig. 3b). Reduction osteotomy is one procedure to match the value of HKA angle and the true alignment (Fig. 3c). In this technique, HKA angle is zero degree, and TM and TA coincide. For figures, please contact authors directly.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 161 - 161
1 Dec 2013
Paniogue T Reinitz S Van Citters D
Full Access

In vivo oxidative degradation in ultra-high molecular weight polyethylene (UHMWPE) has gained significant attention in recent years, especially with the discovery of unanticipated oxidation in retrieved highly cross-linked bearings. While significant attention has been paid to mechanical property changes caused by oxidation, there has been little focus on understanding how wear rates are affected by these in vivo changes. Recent work has demonstrated the possibility of machining wear pins from retrieved UHMWPE bearings, but leveling of the pins removed the in vivo articular surface.[1] The goal of this study is to determine whether wear pins can be produced utilizing the native articular surface. Three materials were used for this study: a short-duration retrieved mobile-bearing conforming tibial insert with minimal oxidation (non-oxidized); a shelf-aged, oxidized, non-conforming fixed bearing tibial insert (oxidized); and standard NIST 1050 bar stock (NIST). Utilizing both conforming and non-conforming devices tests the technique over a range of articular curvatures, while testing a highly oxidized material tests the feasibility of maintaining the native surface when machining wear pins with compromised mechanical properties. FTIR analysis was performed at the articular surface of the devices near where the pins were taken, using ketone peak height as an indicator of oxidation. Wear rates were determined using a six station AMTI OrthoPod with an applied load of 100 N in multidirectional motion for a total of 2 million cycles. The oxidized material had a surface ketone level of 0.26, the non-oxidized device had a ketone level of 0.05, and the NIST sample had a ketone level less than 0.01. Two pins of each material were machined to ¼″ diameter with a length of the through thickness of the tibial inserts; soak controls were also produced. Figure 1 shows mass loss data for all six pins tested. Wear rates between the two pins of each group were fairly repeatable, and the wear rates of the different groups could be easily differentiated. The pins machined from NIST bar stock showed the best match-up, but pins machined from retrieved devices also showed good repeatability, with the non-conforming device showing better results than the conforming device. The ability to produce repeatable wear results with pins machined from in vivo devices is an important step in understanding how the wear rate changes over time in vivo. By maintaining the native articular surface, this test will give a more true representation of the in-vivo wear rate. This method will enable future investigations into how wear rates are affected by oxidation, absorbed chemical species, or other changes that occur in vivo


Orthopaedic Proceedings
Vol. 85-B, Issue SUPP_II | Pages 105 - 105
1 Feb 2003
Hussain SA Lam F Slack R Arya A Compson J
Full Access

Certain cases of patello-femoral maltracking can lead to articular surface wear. Though most can be treated non-operatively, where there is increasing wear surgical intervention may be necessary. Patellar tracking is difficult to assess and though several different types of maltracking or loading have been described, each case warrants precise assessment of the wear patterns. Without this knowledge a logical approach to realignment surgery is impossible. 60 consecutive cases (age range 18–50 years) presenting with anterior knee pain were arthroscoped over a 4 year period. These patients all had been selected with either patellar instability or surface wear indicated either clinically, a positive radiograph, bone scan or MRI. All patients were arthroscoped through standard anterolateral and antero-medial portals and also a superolateral and occasionally a supero-medial approach. The areas of articular damage were mapped on diagrams and recorded photographically. Patella views were taken in flexion and extension, and on passively stretching the patella medially and laterally. We found 6 distinct patterns of wear which appear to indicate 6 different maltracking abnormalities. The largest group, 46 patients, consisted of lateral trackers, with 21 patients demonstrating medial facet and lateral femoral condylar wear. Assessment of the articular surface of the patello-femoral groove from inferior portals is highly misleading and superior portals are needed for proper assessment. Medial facet wear can occur in lateral instability or medial compression. Lateral maltracking at engagement or distally are the commonest patterns


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 425 - 425
1 Apr 2004
Harman M Banks S Hodge W
Full Access

Both backside and articular surface wear have been linked to osteolysis after total knee arthroplasty (TKA). Prostheses with cementless fixation, screw holes in high load regions, and thin polyethylene are susceptible to backside wear. Factors associated with articular wear are similarly well defined. Micromotion at the modular polyethylene interface has been reported for many prostheses, but the relevance of such data compared to articular motions and wear are difficult to appreciate. This study compares in vivo motions and wear occurring at the backside and articular surfaces after TKA. Contemporary PCL-retaining prostheses from one manufacturer were implanted by one surgeon using cement fixation. The polyethylene inserts were > 6mm thick with a full peripheral rim capture and anterior wire locking mechanism. Femoral condylar motions were measured in 20 knees using fluoroscopic analysis during stair and gait activities. All patients had good to excellent clinical outcomes at one year follow-up. Articular and backside surface damage was evaluated on 32 polyethylene inserts retrieved after 27 months (1 to 71) months in-situ for infection (9), autopsy (6), patellar resurfacing (4), patellar loosening (4), tibial loosening (3), osteolysis (2), and other (4). Femoral condylar translation over the polyethylene articular surface ranged from 5-10 mm, which is substantially larger than the reported 50-500 micron range of backside interface micromotion measured in vitro. Damage covered < 33% of the backside surface and appeared as a cast impression of the opposed metal tibial component without scratches associated with micromotion. In contrast, damage consisting predominantly of scratching, burnishing and tractive striations covered 46% of the articular surface. Different locking mechanisms for modular polyethylene inserts result in different degrees of backside wear. No significant backside wear was observed these retrieved inserts with a wire-supplemented peripheral capture. Given the abrasive wear mechanisms and particulate debris shed during femoral condylar sliding, efforts to control motions at the articular surface appear warranted


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 93 - 93
1 Jun 2012
Hirakawa M Kondo M Tomari K Sato M Itokawa T Kaku N Ikeda S Hara K Tsumura H
Full Access

Restoration of joint line in total knee arthroplasty (TKA) is important for kinematics of knee and ligamentous balance. Especially in revision TKA, it may be difficult to identify the joint line. The aim of this study is to define the relationship between epicondyles and articular surface using CT based three-dimensional digital templating sofware $“Athena” (Soft Cube, Osaka, Japan). 137 knees with osteoarthritis, all caces were grade 2 or lower in Kellgren-Lawrence index, were investigated. Perpendicular lines were dropped from the prominences of the medial and lateral femoral epicondyles to the most distal points of articular surfaces and distances of the lines were measured on the axial and coronal planes. The femoral width was measured as the distance between medial and lateral epicondyles. Each of the distance described above was converted to a ratio by dividing by the femoral width. On the axial plane, the average distance from epicondyles to the posterior articular surfaces were 29.4±2.2mm on the medial side and 21.2±2.3mm on the lateral side. The average of the femoral width was 75.2±4.1mm. On coronal plane, the average distance from epicondyles to the distal articular surfaces were 25.2±2.8mm on the medial side and 21.4±2.5mm on the lateral side. The ratio for the distance from epicondyles to the distal and posterior joint line compared to femoral width was 0.39±0.02, 0.28±0.03, 0.33±0.03 and 0.28±0.03. The distance from epicondyles to the distal and posterior joint line correlates with the femoral width of the distal femur. This information can be useful in determining appropriate joint line


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 604 - 604
1 Dec 2013
Zumbrunn T Varadarajan KM Rubash HE Li G Muratoglu O
Full Access

INTRODUCTION. Contemporary PCL sacrificing Total Knee Arthroplasty (TKA) implants (CS) consist of symmetric medial and lateral tibial articular surfaces with high anterior lips designed to substitute for the stability of the native PCL. However, designs vary significantly across implant systems in the level of anteroposterior constraint provided. Therefore, the goal of this study was to investigate kinematics of two CS designs with substantially different constraint levels. The hypothesis was that dynamic knee simulations could show the effect of implant constraint on kinematics of CS implants. METHODS. LifeModeler KneeSIM software was used to analyze contemporary CS TKA (X) with a symmetric and highly dished tibia and contemporary CS TKA (Y) with a symmetric tibia having flat sections bounded by high anterior and posterior lips, during simulated deep knee bend and chair sit. The flat sections of CS-Y implant are designed to allow freedom prior to motion restriction by the implant lips. Components were mounted on an average knee model created from Magnetic Resonance Imaging (MRI) data of 40 normal knees. Relevant ligament/tendon insertions were obtained from the MRI based 3D models and tissue properties were based on literature values. The condyle center motions relative to the tibia were used to compare the different implant designs. In vivo knee kinematics of healthy subjects from published literature was used for reference. RESULTS. Prior publications on in vivo kinematics of healthy knees showed that normal knee motion is characterized by an overall medial pivot. This includes greater, consistent posterior rollback of the lateral condyle than medial (Fig 3). In contrast, CS implant X showed symmetric motion including paradoxical anterior sliding until 120° flexion. This caused a more anterior location for both femoral condyles in flexion as opposed to the posterior location seen in healthy knees. CS implant Y with flat sections showed even greater anterior sliding than CS-X. These trends were seen for chair sit activity as well. Thus, while CS-X showed less paradoxical sliding, both implants suffered from kinematic deficits due to absence of the PCL (Fig. 1 and Fig. 2). CONCLUSION. The two CS implants showed different kinematic performance confirming the hypothesis that implant design affects kinematics of CS TKA. Absence of the PCL in contemporary CS implants resulted in kinematic deficits. In particular, a symmetric implant with flat sections connecting the anteroposterior implant lips showed excessive paradoxical anterior sliding. These data showed both the need and opportunity for novel designs to address the limitations of contemporary CS implants


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 134 - 134
1 May 2016
Esposito C Liu T Burket J Wright T
Full Access

Introduction. Highly crosslinked ultrahigh-molecular-weight polyethylene (XLPE) reduces wear and osteolysis in total hip arthroplasty, but it is unclear if XLPE will provide the same clinical benefit in total knee arthroplasty (TKA). Adhesive and abrasive wear generally dominate in polyethylene acetabular components, whereas fatigue wear is an important wear mechanism in polyethylene TKA tibial inserts. The wear resistance of XLPE depends on the crosslink density of the material, which may decrease during in vivo mechanical loading, leading to more wear and increased oxidation. To examine this possibility, we measured crosslink density and oxidation levels in loaded and unloaded locations of retrieved tibial inserts to evaluate the short-term performance of XLPE material in TKA. Materials and Methods. Forty retrieved XLPE tibial inserts (23 remelted, 17 annealed) retrieved after a mean time of 18 ± 14 months were visibly inspected to identify loaded (burnished) and unloaded (unburnished) locations on the plateaus of each insert using a previously published damage mapping method. For each insert, four cubes (3 mm3) were cut from loaded and unloaded surface and subsurface locations (Fig. 1). Swell ratio testing was done according to ASTM F2214 to calculate crosslink density of the cubes. With a microtome, 200 μm sections were taken adjacent to the cubes and oxidation was assessed with Fourier transform infrared spectroscopy following ASTM F2102 (Fig. 2). Surface oxidation was measured in the sections adjacent the surface cubes and subsurface oxidation was measured in sections adjacent to the subsurface cubes. The effects of location (surface vs. subsurface in the loaded and unloaded regions) and thermal treatment (annealed vs. remelted) on crosslink density and oxidation were assessed with repeated measures generalized estimating equations (GEEs), with the implant treated as the repeated factor. Results are presented as means and 95% confidence intervals and the level of significance was α=0.05. Results. Crosslink density was associated with location within the polyethylene tibial inserts (p<0.001), while oxidation was associated with both location (p<0.001) and heat treatment (p=0.003). The loaded surface (location 1 in Fig. 1) had 13% lower crosslink density than all other locations (p<0.001 for each), and greater oxidation than all other locations (Fig. 3). Specifically, oxidation of the loaded surface was 0.29[0.17,0.40] greater (two times greater) than that of the unloaded surface (p < 0.001), whereas subsurface areas of loaded and unloaded regions differed by only 0.03[0.00,0.07] (p<0.022). Additionally, surface oxidation was over 7-fold greater than subsurface oxidation in the loaded region (difference: 0.56[0.44,0.68], p<0.001). Annealed XLPE had 2-fold greater oxidation than remelted XLPE (difference 0.159, 95% CI = 0.045, 0.126), and this was independent of location within the inserts. Conclusions. In vivo loading of XLPE decreased the crosslink density and increased the oxidation in areas that underwent wear and deformation at the articular surface of TKA inserts. Nonetheless, in these short term retrievals, no clinical complications were attributed to the change in material properties. However, if crosslink density continues to decrease with load over time, XLPE may not provide a clinical advantage over conventional polyethylene in TKA


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 25 - 25
1 May 2016
Seo J Yoo O In Y Kang B Sun D Kim Y
Full Access

Introduction

Knee joint should be aligned for reconstruction of the function in Total Knee Replacement(TKR). Although a surgeon try to correct the alignment of a knee joint, sometimes varus/valgus alignment has been tried in order to reconstruct function of knee joint. As a result, the varus or valgus alignment affects to ligaments and soft tissue, and the contact condition is changed between femoral component and tibial insert. One of important factor, wear characteristics of an implant can be changed due to the contact condition. In this study, we performed static contact tests from extension to flexion in varus and valgus to define the effect to contact condition when the alignment is varus or valgus.

Methods

LOSPA TKR femoral component #6 and Tibial insert #5 manufactured by Corentec Co., Ltd. were used as test specimens. The tests have performed with adapting ASTM F2777–10 ‘Standard Test Method for Evaluating Knee Bearing (Tibial Insert) Endurance and Deformation under High Flexion’. The test set like as Fig. 1. The load is applied at 7:3 ratio of lateral-medial by adapting gait analysis. The 5° of jig is used to compare the result in neutral, varus and valgus. The fuji films were used in tests were scanned, and the results were analyzed the compressed area and contact stress as angles of flexion in neutral stance and varus/valgus from scanning. The tests were performed 5 times per each for a reliability.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 376 - 377
1 Jul 2008
Harrold F Park-Wesley F Abboud R Wigderowitz C
Full Access

Introduction: Successful shoulder arthroplasty is based on restoration of the individual’s proximal humeral morphology with a precise osteotomy of the humeral head at the level of the anatomical neck. The objective of this study was to determine the geometry of the articular portion of the humeral head in contact with the glenoid in the neutral position and compare the orientation to the geometry of the humeral head determined using the cartilage/calcar interface of the anatomical neck.

Methods: An intact rotator cuff and joint capsule were exposed for six cadaveric full arms. Precision perspex reference cubes were attached to the greater tuberosity of the humerus and to the coracoid process of the scapula on each specimen. Each shoulder was mounted in a custom built jig with the arm fixed in the neutral position and a Microscribe 3D-X digitizer used to digitize three faces of each precision cube. The shoulder joint was then disarticulated and both the humerus and scapula re-mounted on the same jig, independently. The cube faces were re-digitized and relevant points, lines and surfaces were identified and digitized on each humerus and scapula. The humeri were then scanned using a high precision surface laser scanner.

The data collected from both digitizing tools were merged into the same coordinate system and graphically represented. Paired Student’s t-tests were used to compare the inclination and retroversion angles for the two techniques.

Results and discussion: The study found a significant difference in inclination (p less than 0.02) and no difference in retroversion (p equal to 0.75) when the glenoid position was used to calculate humeral head orientation (Inclination: Mean 11.5 deg., StD. 11.2 deg.; Retroversion: Mean 20.5 deg., StD. 6.6 deg.) when compared to using the cartilage/calcar interface (Inclination: Mean 134.1 deg., StD. 1.9 deg.; Retroversion Mean 21.7 deg., StD. 13.9 deg.).

Small deviations in the recovery of head orientation in shoulder arthroplasty may impact on the longevity of an implant. The differences in inclination and retroversion noted in this study may alter the load on the glenoid and/or rotator cuff mechanism in joint replacement. Further research is necessary.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_9 | Pages 28 - 28
17 Apr 2023
Jimenez-Cruz D Dubey M Board T Williams S
Full Access

Hip joint biomechanics can be altered by abnormal morphology of the acetabulum and/or femur. This may affect load distribution and contact stresses on the articular surfaces, hence, leading to damage and degradation of the tissue. Experimental hip joint simulators have been used to assess tribology of total hip replacements and recently methods further developed to assess the natural hip joint mechanics. The aim of this study was to evaluate articular surfaces of human cadaveric joints following prolonged experimental simulation under a standard gait cycle. Four cadaveric male right hips (mean age = 62 years) were dissected, the joint disarticulated and capsule removed. The acetabulum and femoral head were mounted in an anatomical hip simulator (Simulation Solutions, UK). A simplified twin peak gait cycle (peak load of 3kN) was applied. Hips were submerged in Ringers solution (0.04% sodium azide) and testing conducted at 1 Hertz for 32 hours (115,200 cycles). Soft tissue degradation was recorded using photogrammetry at intervals throughout testing. All four hips were successfully tested. Prior to simulation, two samples exhibited articular surface degradation and one had a minor scalpel cut and a small area of cartilage delamination. The pre-simulation damage got slightly worse as the simulation continued but no new areas of damage were detected upon inspection. The samples without surface degradation, showed no damage during testing and the labral sealing effect was more obvious in these samples. The fact that no new areas of damage were detected after long simulations, indicates that the loading conditions and positioning of the sample were appropriate, so the simulation can be used as a control to compare mechanical degradation of the natural hip when provoked abnormal conditions or labral tissue repairs are simulated


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_7 | Pages 19 - 19
1 Jul 2022
Sweed T Boutefnouchet T Lim Z Amerasekera S Choudhary S Ashraf T
Full Access

Abstract. Introduction. There are several imaging-based measurements for patello-femoral height. Available methods rely predominantly on sagittal images. The latter can be misleading with sagittal oblique slices and when the patella is tilted and/or chronically subluxed. In this study we describe a simple method of patellar height measurement using axial MRI overlap. Materials and methods. A retrospective observational analysis of 97 knees from 251 patients was conducted. Cases were selected following the exclusion of scans with fractures, massive effusion, patello-femoral pathology. Axial patello-trochlear overlap (APTO) was measured on the axial MRI images as follows: (1) Patellar length (P): expressed as the number of axial images showing patellar articular surface (2) Trochlear overlap (T): the number of axial images showing overlap between patellar articular surface and articular surface of lateral trochlea. APTO is the ratio T/P. All measurements were carried out independently and on two separate occasions by 6 raters. As a control conventional patello-trochlear index were measured for all patients by a senior musculoskeletal radiologist. Results. The mean APTO value was 36.7 (range 14.2 to 66.6; SD 11.4). There was a positive correlation with patello-trochlear index (Pearson correlation coefficient: 0.76, P < 0.001). Intra-observer reliability was good (ICC: 0.66 95 CI 0.54, 0.76, P < 0.001). Inter-observer reliability was fair (ICC: 0.51, 95 CI 0.41, 0.6, P < 0.001). Conclusion. In the present proof of concept study APTO was a reliable measurement of patellar height and correlated with patella-trochlear indices. The method described can prove valuable in overcoming issues with sagittal image measurements


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 102 - 102
1 Apr 2019
Mani S Wernle J
Full Access

Introduction. A search of the literature indicates several constrained total knee arthroplasty (TKA) systems are at risk for articular surface lockdown bolts backing out. The backing out of a lockdown bolt may lead to an unstable and/or painful knee and may necessitate revision. Upon backing out, the bolt may damage implant components and surrounding tissues. To date, studies in the literature have not simulated or replicated loosening of bolts in TKA. Therefore, the objectives of this study were to 1) develop a set of physiological loading parameters that challenge bolted articular surfaces; 2) evaluate whether significant bolt torque is lost during application of this loading to a CCK device with a bolt as a secondary locking mechanism. Materials and Methods. Physical test parameters to loosen lockdown bolts were developed based on loading experienced during activities of daily living. Sinusoidal waveforms and timing were used to simulate worst case walking gait conditions. Compared to data from everyday activities in instrumented TKR patients, anterior posterior loads and internal/external torques exceeding the absolute maximums observed were selected. To transfer more shear and torsion to the joint interface, compressive load lower than typically reported for walking gait was used. Frequency was representative of walking gait motion. The offset in torsional waveform enables a ratcheting motion to drive a loose bolt out of the joint: during external femoral rotation of a left knee, reduced compressive load and posterior directed femoral loading on a CCK spine creates a potential articular surface lift-off. The lift-off may grab the underside of the front bolt shoulder while external (CCW) rotation loosens the bolt. These loading conditions exist during toe-off of walking gait. Two CCK devices were evaluated to capture potential difference in performance: a medium articular surface combination and a smaller articular surface combination. Testing was performed on a load frame capable of rotation and vertical / horizontal translation. Results. No bolts completely loosened when fully tightened. However, average loss in bolt torque of 39.3% on the medium and 21.5% on the smaller articular surface was observed. Loading led to reorientation of the articular surfaces verified by markings on the components. Additional constructs that were under-tightened intentionally to one-quarter of target torque value lost all bolt torque and completely backed out. Discussion and Significance. The backing out of lockdown bolts in TKA has been reported in the literature but not replicated in-vitro. A challenging, physiologically relevant set of loading parameters was developed and applied to a CCK device with an articular surface lockdown bolt. Upon loading, the bolts experienced statistically significant loss of bolt torque which may be attributed to articular surface reorientation. Selected loading parameters led to complete bolt back-out in under-tightened constructs


Bone & Joint Open
Vol. 5, Issue 5 | Pages 394 - 400
15 May 2024
Nishi M Atsumi T Yoshikawa Y Okano I Nakanishi R Watanabe M Usui Y Kudo Y

Aims. The localization of necrotic areas has been reported to impact the prognosis and treatment strategy for osteonecrosis of the femoral head (ONFH). Anteroposterior localization of the necrotic area after a femoral neck fracture (FNF) has not been properly investigated. We hypothesize that the change of the weight loading direction on the femoral head due to residual posterior tilt caused by malunited FNF may affect the location of ONFH. We investigate the relationship between the posterior tilt angle (PTA) and anteroposterior localization of osteonecrosis using lateral hip radiographs. Methods. Patients aged younger than 55 years diagnosed with ONFH after FNF were retrospectively reviewed. Overall, 65 hips (38 males and 27 females; mean age 32.6 years (SD 12.2)) met the inclusion criteria. Patients with stage 1 or 4 ONFH, as per the Association Research Circulation Osseous classification, were excluded. The ratios of anterior and posterior viable areas and necrotic areas of the femoral head to the articular surface were calculated by setting the femoral head centre as the reference point. The PTA was measured using Palm’s method. The association between the PTA and viable or necrotic areas of the femoral head was assessed using Spearman’s rank correlation analysis (median PTA 6.0° (interquartile range 3 to 11.5)). Results. We identified a negative correlation between PTA and anterior viable areas (rho −0.477; p = 0.001), and no correlation between PTA and necrotic (rho 0.229; p = 0.067) or posterior viable areas (rho 0.204; p = 0.132). Conclusion. Our results suggest that residual posterior tilt after FNF could affect the anteroposterior localization of necrosis. Cite this article: Bone Jt Open 2024;5(5):394–400


Bone & Joint Research
Vol. 8, Issue 5 | Pages 207 - 215
1 May 2019
Key S Scott G Stammers JG Freeman MAR Pinskerova V Field RE Skinner J Banks SA

Objectives. The medially spherical GMK Sphere (Medacta International AG, Castel San Pietro, Switzerland) total knee arthroplasty (TKA) was previously shown to accommodate lateral rollback while pivoting around a stable medial compartment, aiming to replicate native knee kinematics in which some coronal laxity, especially laterally, is also present. We assess coronal plane kinematics of the GMK Sphere and explore the occurrence and pattern of articular separation during static and dynamic activities. Methods. Using pulsed fluoroscopy and image matching, the coronal kinematics and articular surface separation of 16 well-functioning TKAs were studied during weight-bearing and non-weight-bearing, static, and dynamic activities. The closest distances between the modelled articular surfaces were examined with respect to knee position, and proportions of joint poses exhibiting separation were computed. Results. Overall, 1717 joint poses were analyzed. At a 1.0 mm detection threshold, 37 instances of surface separation were observed in the lateral compartment and four medially (p < 0.001). Separation was activity-dependent, both laterally and medially (p < 0.001), occurring more commonly during static deep flexion in the lateral compartment, and during static rotation in the medial compartment. Lateral separation occurred more frequently than medial during kneeling (7/14 lateral vs 1/14 medial; p = 0.031) and stepping (20/1022 lateral vs 0/1022 medial; p < 0.001). Separation varied significantly between individuals during dynamic activities. Conclusion. No consistent association between closest distances of the articular surfaces and knee position was found during any activity. Lift-off was infrequent and depended on the activity performed and the individual knee. Lateral separation was consistent with the design rationale. Medial lift-off was rare and mostly in non-weight-bearing activities. Cite this article: S. Key, G. Scott, J. G. Stammers, M. A. R. Freeman†, V. Pinskerova, R. E. Field, J. Skinner, S. A. Banks. Does lateral lift-off occur in static and dynamic activity in a medially spherical total knee arthroplasty? A pulsed-fluoroscopic investigation. Bone Joint Res 2019;8:207–215. DOI: 10.1302/2046-3758.85.BJR-2018-0237.R1


Bone & Joint Open
Vol. 3, Issue 5 | Pages 390 - 397
1 May 2022
Hiranaka T Suda Y Saitoh A Tanaka A Arimoto A Koide M Fujishiro T Okamoto K

The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA, the tibial cutting surface is decided parallel to the femoral cutting surface (or trial component) with in-line traction. These approaches are categorized as unrestricted KA because there is no consideration of leg alignment or component orientation. Restricted KA is an approach where the periarthritic joint surface is replicated within a safe range, due to concerns about extreme alignments that have been considered ‘alignment outliers’ in the neutral mechanical alignment approach. More recently, functional alignment and inverse kinematic alignment have been advocated, where bone cuts are made following intraoperative planning, using intraoperative measurements acquired with computer assistance to fulfill good coordination of soft-tissue balance and alignment. The KA-TKA approach aims to restore the patients’ own harmony of three knee elements (morphology, soft-tissue balance, and alignment) and eventually the patients’ own kinematics. The respective approaches start from different points corresponding to one of the elements, yet each aim for the same goal, although the existing implants and techniques have not yet perfectly fulfilled that goal


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_8 | Pages 1 - 1
1 Aug 2020
Culliton K Speirs A Beaulé P
Full Access

The avascular nature of articular cartilage relies on diffusion pathways to obtain essential nutrients and molecules for cellular activity. Understanding these transport pathways is essential to maintaining and improving the health of articular cartilage and ultimately synovial joints. Several studies have shown that joint articulation is associated with fluid and solute uptake although it remains unclear what role sliding motion independently plays. This study investigates the role of sliding with a non-stationary contact area on the uptake of small molecular weight tracers into articular cartilage. Ten-millimeter diameter cartilage-bone plugs were obtained from porcine knee joints and sealed into purpose made diffusion chambers. The chambers were designed to eliminate diffusion from the radial edge and only allow diffusion through the articular surface. The bone side of the chamber was filled with PBS to maintain tissue hydration while the cartilage side was filled with 0.01mg/ml fluorescein sodium salt (FNa) prepared using PBS. Sliding loads with a non-stationary contact area were applied across the articular surface by a custom apparatus using a 4.5 mm diameter spherical indenter. A moving contact area was chosen to represent physiological joint motions. Reciprocal sliding was maintained at a rate of 5 mm/s for 2 and 4 hours. Control samples were subject to passive diffusion for 0, 4, and 88 hours. After diffusion tests, samples were snap frozen and 20 µm cross-sectional cuts were taken perpendicular to the sliding direction. Samples were imaged using a Zeiss AxioImager M2 epifluorescent microscope under 5× magnification with a filter for FNa. Intensity profiles were mapped from the articular surface to the subchondral bone. Unloaded control samples demonstrated minimal solute uptake at 4 hours penetrating less than 5% of the total cartilage depth. By 88 hours solute penetration had reached the subchondral bone although there was minimal accumulation within the cartilage matrix indicated by the relatively low intensity profile values. Samples that had been subjected to reciprocal sliding demonstrated accelerated penetration and solute accumulation compared to unloaded samples. After 1 hour of reciprocal sliding, the solute had reached 40% of the cartilage depth, this increased to approximately 80% at 4 hours, with much higher intensities compared to unloaded controls. Sliding motion plays an important role in the uptake of solutes into the cartilage matrix. Maintaining joint motion both post injury and in the arthritic process is a critical component of cartilage nutrition. Samples that had been subject to reciprocal sliding demonstrated accelerated solute penetration and accumulation in the cartilage matrix, exceeding steady state concentrations achieved by passive diffusion


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_11 | Pages 75 - 75
1 Dec 2020
Burkhard B Schopper C Ciric D Mischler D Gueorguiev B Varga P
Full Access

Proximal humerus fractures (PHF) are the third most common fractures in the elderly. Treatment of complex PHF has remained challenging with mechanical failure rates ranging up to 35% even when state-of-the-art locked plates are used. Secondary (post-operative) screw perforation through the articular surface of the humeral head is the most frequent mechanical failure mode, with rates up to 23%. Besides other known risk factors, such as non-anatomical reduction and lack of medial cortical support, in-adverse intraoperative perforation of the articular surfaces during pilot hole drilling (overdrilling) may increase the risk of secondary screw perforation. Overdrilling often occurs during surgical treatment of osteoporotic PHF due to minimal tactile feedback; however, the awareness in the surgical community is low and the consequences on the fixation stability have remained unproved. Therefore, the aim of this study was to evaluate biomechanically whether overdrilling would increase the risk of cyclic screw perforation failure in unstable PHF. A highly unstable malreduced 3-part fracture was simulated by osteotomizing 9 pairs of fresh-frozen human cadaveric proximal humeri from elderly donors (73.7 ± 13.0 ys, f/m: 3/6). The fragments were fixed with a locking plate (PHILOS, DePuy Synthes, Switzerland) using six proximal screws, with their lengths selected to ensure 6 mm tip-to-joint distance. The pairs were randomized into two treatment groups, one with all pilot holes accurately predrilled (APD) and another one with the boreholes of the two calcar screws overdrilled (COD). The constructs were tested under progressively increasing cyclic loading to failure at 4 Hz using a previously developed setup and protocol. Starting from 50 N, the peak load was increased by 0.05 N/cycle. The event of initial screw loosening was defined by the abrupt increase of the displacement at valley load, following its initial linear behavior. Perforation failure was defined by the first screw penetrating the joint surface, touching the artificial glenoid component and stopping the test via electrical contact. Bone mineral density (range: 63.8 – 196.2 mgHA/cm3) was not significantly different between the groups. Initial screw loosening occurred at a significantly lower number of cycles in the COD group (10,310 ± 3,575) compared to the APD group (12,409 ± 4,569), p = 0.006. Number of cycles to screw perforation was significantly lower for the COD versus APD specimens (20,173 ± 5,851 and 24,311 ± 6,318, respectively), p = 0.019. Failure mode was varus collapse combined with lateral-inferior translation of the humeral head. The first screw perforating the articular surface was one of the calcar screws in all but one specimen. Besides risk factors such as fracture complexity and osteoporosis, inadequate surgical technique is a crucial contributor to high failure rates in locked plating of complex PHF. This study shows for the first time that overdrilling of pilot holes can significantly increase the risk of secondary screw perforation. Study limitations include the fracture model and loading method. While the findings require clinical corroboration, raising the awareness of the surgical community towards this largely neglected risk source, together with development of devices to avoid overdrilling, are expected to help improve the treatment outcomes


Bone & Joint Open
Vol. 3, Issue 6 | Pages 495 - 501
14 Jun 2022
Keohane D Sheridan GA Masterson E

Aims. Total knee arthroplasty (TKA) is a common and safe orthopaedic procedure. Zimmer Biomet's NexGen is the second most popular brand of implant used in the UK. The primary cause of revision after the first year is aseptic loosening. We present our experience of using this implant, with significant concerns around its performance with regards early aseptic loosening of the tibial component. Methods. A retrospective, single-surgeon review was carried out of all of the NexGen Legacy Posterior Stabilized (LPS) TKAs performed in this institute. The specific model used for the index procedures was the NexGen Complete Knee System (Legacy Knee-Posterior Stabilized LPS-Flex Articular Surface, LPS-Flex Femoral Component Option, and Stemmed Nonaugmentable Tibial Component Option). Results. Between 2013 and 2016, 352 NexGen TKAs were carried out on 331 patients. A total of 62 TKAs have been revised to date, giving an all-cause revision rate of 17.6% at a minimum of five years. Three of these revisions were due to infection. Overall, 59 of the revisions were performed for aseptic loosening (16.7%) of the tibial component. The tibial component was removed intraoperatively without instrumentation due to significant tibial debonding between the implant-cement interface. Conclusion. While overall, we believe that early aseptic loosening is multi-factorial in nature, the significantly high aseptic revision rate, as seen by an experienced fellowship-trained arthroplasty surgeon, has led us to believe that there is a fundamental issue with this NexGen implant design. Continued implant surveillance and rigorous review across all regions using this particular implant is warranted based on the concerning findings described here. Cite this article: Bone Jt Open 2022;3(6):495–501