Advertisement for orthosearch.org.uk
Results 1 - 9 of 9
Results per page:
Bone & Joint Open
Vol. 4, Issue 10 | Pages 808 - 816
24 Oct 2023
Scott CEH Snowden GT Cawley W Bell KR MacDonald DJ Macpherson GJ Yapp LZ Clement ND

Aims. This prospective study reports longitudinal, within-patient, patient-reported outcome measures (PROMs) over a 15-year period following cemented single radius total knee arthroplasty (TKA). Secondary aims included reporting PROMs trajectory, 15-year implant survival, and patient attrition from follow-up. Methods. From 2006 to 2007, 462 consecutive cemented cruciate-retaining Triathlon TKAs were implanted in 426 patients (mean age 69 years (21 to 89); 290 (62.7%) female). PROMs (12-item Short Form Survey (SF-12), Oxford Knee Score (OKS), and satisfaction) were assessed preoperatively and at one, five, ten, and 15 years. Kaplan-Meier survival and univariate analysis were performed. Results. At 15 years, 28 patients were lost to follow-up (6.1%) and 221 patients (51.9%) had died, with the mean age of the remaining cohort reducing by four years. PROMs response rates among surviving patients were: one-year 63%; five-year 72%; ten-year 94%; and 15-year 59%. OKS and SF-12 scores changed significantly over 15 years (p < 0.001). The mean improvement in OKS was 18.8 (95% confidence (CI) 16.7 to 19.0) at one year. OKS peaked at five years (median 43 years) declining thereafter (p < 0.001), though at 15 years it remained 17.5 better than preoperatively. Age and sex did not alter this trajectory. A quarter of patients experienced a clinically significant decline (≥ 7) in OKS from five to ten years and from ten to 15 years. The SF-12 physical component score displayed a similar trajectory, peaking at one year (p < 0.001). Patient satisfaction was 88% at one, five, and ten years, and 94% at 15 years. In all, 15-year Kaplan-Meier survival was 97.6% (95% CI 96.0% to 99.2%) for any revision, and 98.9% (95% CI 97.9% to 99.9%) for aseptic revision. Conclusion. Improvements in PROMs were significant and maintained following single radius TKA, with OKS peaking at five years, and generic physical health peaking at one year. Patient satisfaction remained high at 15 years, at which point 2.4% had been revised. Cite this article: Bone Jt Open 2023;4(10):808–816


Bone & Joint Research
Vol. 1, Issue 12 | Pages 315 - 323
1 Dec 2012
Molt M Ljung P Toksvig-Larsen S

Objectives

The objective of this study was to compare the early migration characteristics and functional outcome of the Triathlon cemented knee prosthesis with its predecessor, the Duracon cemented knee prosthesis (both Stryker).

Methods

A total 60 patients were prospectively randomised and tibial component migration was measured by radiostereometric analysis (RSA) at three months, one year and two years; clinical outcome was measured by the American Knee Society score and the Knee Osteoarthritis and Injury Outcome Score.


Bone & Joint Open
Vol. 5, Issue 10 | Pages 911 - 919
21 Oct 2024
Clement N MacDonald DJ Hamilton DF Gaston P

Aims

The aims were to assess whether joint-specific outcome after total knee arthroplasty (TKA) was influenced by implant design over a 12-year follow-up period, and whether patient-related factors were associated with loss to follow-up and mortality risk.

Methods

Long-term follow-up of a randomized controlled trial was undertaken. A total of 212 patients were allocated a Triathlon or a Kinemax TKA. Patients were assessed preoperatively, and one, three, eight, and 12 years postoperatively using the Oxford Knee Score (OKS). Reasons for patient lost to follow-up, mortality, and revision were recorded.


Bone & Joint Open
Vol. 4, Issue 1 | Pages 13 - 18
5 Jan 2023
Walgrave S Oussedik S

Abstract

Robotic-assisted total knee arthroplasty (TKA) has proven higher accuracy, fewer alignment outliers, and improved short-term clinical outcomes when compared to conventional TKA. However, evidence of cost-effectiveness and individual superiority of one system over another is the subject of further research. Despite its growing adoption rate, published results are still limited and comparative studies are scarce. This review compares characteristics and performance of five currently available systems, focusing on the information and feedback each system provides to the surgeon, what the systems allow the surgeon to modify during the operation, and how each system then aids execution of the surgical plan.

Cite this article: Bone Jt Open 2023;4(1):13–18.


Bone & Joint Open
Vol. 4, Issue 6 | Pages 432 - 441
5 Jun 2023
Kahlenberg CA Berube EE Xiang W Manzi JE Jahandar H Chalmers BP Cross MB Mayman DJ Wright TM Westrich GH Imhauser CW Sculco PK

Aims

Mid-level constraint designs for total knee arthroplasty (TKA) are intended to reduce coronal plane laxity. Our aims were to compare kinematics and ligament forces of the Zimmer Biomet Persona posterior-stabilized (PS) and mid-level designs in the coronal, sagittal, and axial planes under loads simulating clinical exams of the knee in a cadaver model.

Methods

We performed TKA on eight cadaveric knees and loaded them using a robotic manipulator. We tested both PS and mid-level designs under loads simulating clinical exams via applied varus and valgus moments, internal-external (IE) rotation moments, and anteroposterior forces at 0°, 30°, and 90° of flexion. We measured the resulting tibiofemoral angulations and translations. We also quantified the forces carried by the medial and lateral collateral ligaments (MCL/LCL) via serial sectioning of these structures and use of the principle of superposition.


Bone & Joint Research
Vol. 7, Issue 5 | Pages 325 - 326
1 May 2018
Clement ND Deehan DJ


Objectives

Posterior condylar offset (PCO) and posterior tibial slope (PTS) are critical factors in total knee arthroplasty (TKA). A computational simulation was performed to evaluate the biomechanical effect of PCO and PTS on cruciate retaining TKA.

Methods

We generated a subject-specific computational model followed by the development of ± 1 mm, ± 2 mm and ± 3 mm PCO models in the posterior direction, and -3°, 0°, 3° and 6° PTS models with each of the PCO models. Using a validated finite element (FE) model, we investigated the influence of the changes in PCO and PTS on the contact stress in the patellar button and the forces on the posterior cruciate ligament (PCL), patellar tendon and quadriceps muscles under the deep knee-bend loading conditions.


Bone & Joint Research
Vol. 6, Issue 1 | Pages 43 - 51
1 Jan 2017
Nakamura S Tian Y Tanaka Y Kuriyama S Ito H Furu M Matsuda S

Objectives

Little biomechanical information is available about kinematically aligned (KA) total knee arthroplasty (TKA). The purpose of this study was to simulate the kinematics and kinetics after KA TKA and mechanically aligned (MA) TKA with four different limb alignments.

Materials and Methods

Bone models were constructed from one volunteer (normal) and three patients with three different knee deformities (slight, moderate and severe varus). A dynamic musculoskeletal modelling system was used to analyse the kinematics and the tibiofemoral contact force. The contact stress on the tibial insert, and the stress to the resection surface and medial tibial cortex were examined by using finite element analysis.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 64 - 70
1 Jan 2015
Hamilton DF Burnett R Patton JT Howie CR Moran M Simpson AHRW Gaston P

Total knee arthroplasty (TKA) is an established and successful procedure. However, the design of prostheses continues to be modified in an attempt to optimise the functional outcome of the patient.

The aim of this study was to determine if patient outcome after TKA was influenced by the design of the prosthesis used.

A total of 212 patients (mean age 69; 43 to 92; 131 female (62%), 81 male (32%)) were enrolled in a single centre double-blind trial and randomised to receive either a Kinemax (group 1) or a Triathlon (group 2) TKA.

Patients were assessed pre-operatively, at six weeks, six months, one year and three years after surgery. The outcome assessments used were the Oxford Knee Score; range of movement; pain numerical rating scales; lower limb power output; timed functional assessment battery and a satisfaction survey. Data were assessed incorporating change over all assessment time points, using repeated measures analysis of variance longitudinal mixed models. Implant group 2 showed a significantly greater range of movement (p = 0.009), greater lower limb power output (p = 0.026) and reduced report of ‘worst daily pain’ (p = 0.003) over the three years of follow-up. Differences in Oxford Knee Score (p = 0.09), report of ‘average daily pain’ (p = 0.57) and timed functional performance tasks (p = 0.23) did not reach statistical significance. Satisfaction with outcome was significantly better in group 2 (p = 0.001).

These results suggest that patient outcome after TKA can be influenced by the prosthesis used.

Cite this article: Bone Joint J 2015;97-B:64–70.