Aims. Arthroplasty surgery of the knee and hip is performed in two to three million patients annually. Periprosthetic joint infections occur in 4% of these patients. Debridement, antibiotics, and implant retention (DAIR) surgery aimed at cleaning the infected prosthesis often fails, subsequently requiring invasive revision of the complete prosthetic reconstruction. Infection-specific imaging may help to guide DAIR. In this study, we evaluated a bacteria-specific hybrid tracer (. 99m. Tc-UBI. 29-41. -Cy5) and its ability to visualize the bacterial load on
Aims. In computer simulations, the shape of the range of motion (ROM) of a stem with a cylindrical neck design will be a perfect cone. However, many modern stems have rectangular/oval-shaped necks. We hypothesized that the rectangular/oval stem neck will affect the shape of the ROM and the prosthetic impingement. Methods. Total hip arthroplasty (THA) motion while standing and sitting was simulated using a MATLAB model (one stem with a cylindrical neck and one stem with a rectangular neck). The primary predictor was the geometry of the neck (cylindrical vs rectangular) and the main outcome was the shape of ROM based on the prosthetic impingement between the neck and the liner. The secondary outcome was the difference in the ROM provided by each neck geometry and the effect of the pelvic tilt on this ROM. Multiple regression was used to analyze the data. Results. The stem with a rectangular neck has increased internal and external rotation with a quatrefoil cross-section compared to a cone in a cylindrical neck. Modification of the cup orientation and pelvic tilt affected the direction of projection of the cone or quatrefoil shape. The mean increase in internal rotation with a rectangular neck was 3.4° (0° to 7.9°; p < 0.001); for external rotation, it was 2.8° (0.5° to 7.8°; p < 0.001). Conclusion. Our study shows the importance of attention to
Aims. Surgeons and most engineers believe that bone compaction improves implant primary stability without causing undue damage to the bone itself. In this study, we developed a murine distal
United Classification System (UCS) B2 and B3 periprosthetic fractures in total hip arthroplasties (THAs) have been commonly managed with modular tapered stems. No study has evaluated the use of monoblock fluted tapered titanium stems for this indication. This study aimed to evaluate the effects of a monoblock stems on implant survivorship, postoperative outcomes, radiological outcomes, and osseointegration following treatment of THA UCS B2 and B3 periprosthetic fractures. A retrospective review was conducted of all patients who underwent revision THA (rTHA) for periprosthetic UCS B2 and B3 periprosthetic fracture who received a single design monoblock fluted tapered titanium stem at two large, tertiary care, academic hospitals. A total of 72 patients met inclusion and exclusion criteria (68 UCS B2, and four UCS B3 fractures). Primary outcomes of interest were radiological stem subsidence (> 5 mm), radiological osseointegration, and fracture union. Sub-analysis was also done for 46 patients with minimum one-year follow-up.Aims
Methods
Metal-on-metal hip resurfacing (MoM-HR) has seen decreased usage due to safety and longevity concerns. Joint registries have highlighted the risks in females, smaller hips, and hip dysplasia. This study aimed to identify if reported risk factors are linked to revision in a long-term follow-up of MoM-HR performed by a non-designer surgeon. A retrospective review of consecutive MoM hip arthroplasties (MoM-HRAs) using Birmingham Hip Resurfacing was conducted. Data on procedure side, indication, implant sizes and orientation, highest blood cobalt and chromium ion concentrations, and all-cause revision were collected from local and UK National Joint Registry records.Aims
Methods
Aseptic loosening is the most common cause of failure following cemented total knee arthroplasty (TKA), and has been linked to poor cementation technique. We aimed to develop a consensus on the optimal technique for component cementation in TKA. A UK-based, three-round, online modified Delphi Expert Consensus Study was completed focusing on cementation technique in TKA. Experts were identified as having a minimum of five years’ consultant experience in the NHS and fulfilling any one of the following criteria: a ‘high volume’ knee arthroplasty practice (> 150 TKAs per annum) as identified from the National joint Registry of England, Wales, Northern Ireland and the Isle of Man; a senior author of at least five peer reviewed articles related to TKA in the previous five years; a surgeon who is named trainer for a post-certificate of comletion of training fellowship in TKA.Aims
Methods
Aims. Osseointegrated prosthetic limbs allow better mobility than socket-mounted prosthetics for lower limb amputees. Fractures, however, can occur in the residual limb, but they have rarely been reported. Approximately 2% to 3% of amputees with socket-mounted prostheses may fracture within five years. This is the first study which directly addresses the risks and management of periprosthetic osseointegration fractures in amputees. Methods. A retrospective review identified 518 osseointegration procedures which were undertaken in 458 patients between 2010 and 2018 for whom complete medical records were available. Potential risk factors including time since amputation, age at osseointegration, bone density, weight, uni/bilateral implantation and sex were evaluated with multiple logistic regression. The mechanism of injury, technique and implant that was used for fixation of the fracture, pre-osseointegration and post fracture mobility (assessed using the K-level) and the time that the prosthesis was worn for in hours/day were also assessed. Results. There were 22 periprosthetic fractures; they occurred exclusively in the femur: two in the femoral neck, 14 intertrochanteric and six subtrochanteric, representing 4.2% of 518 osseointegration operations and 6.3% of 347
The aims of this study were to determine the incidence and factors for developing periprosthetic joint infection (PJI) following hemiarthroplasty (HA) for hip fracture, and to evaluate treatment outcome and identify factors associated with treatment outcome. A retrospective review was performed of consecutive patients treated for HA PJI at a tertiary referral centre with a mean 4.5 years’ follow-up (1.6 weeks to 12.9 years). Surgeries performed included debridement, antibiotics, and implant retention (DAIR) and single-stage revision. The effect of different factors on developing infection and treatment outcome was determined.Aims
Methods
Unicompartmental knee arthroplasty (UKA) is the preferred treatment for anterior medial knee osteoarthritis (OA) owing to the rapid postoperative recovery. However, the risk factors for UKA failure remain controversial. The clinical data of Oxford mobile-bearing UKAs performed between 2011 and 2017 with a minimum follow-up of five years were retrospectively analyzed. Demographic, surgical, and follow-up data were collected. The Cox proportional hazards model was used to identify the risk factors that contribute to UKA failure. Kaplan-Meier survival was used to compare the effect of the prosthesis position on UKA survival.Aims
Methods
The aims of the study were to report for a cohort aged younger than 40 years: 1) indications for HRA; 2) patient-reported outcomes in terms of the modified Harris Hip Score (HHS); 3) dislocation rate; and 4) revision rate. This retrospective analysis identified 267 hips from 224 patients who underwent an hip resurfacing arthroplasty (HRA) from a single fellowship-trained surgeon using the direct lateral approach between 2007 and 2019. Inclusion criteria was minimum two-year follow-up, and age younger than 40 years. Patients were followed using a prospectively maintained institutional database.Aims
Methods
Cementing in arthroplasty for hip fracture is associated with improved postoperative function, but may have an increased risk of early mortality compared to uncemented fixation. Quantifying this mortality risk is important in providing safe patient care. This study investigated the association between cement use in arthroplasty and mortality at 30 days and one year in patients aged 50 years and over with hip fracture. This retrospective cohort study used linked data from the Australian Hip Fracture Registry and the National Death Index. Descriptive analysis and Kaplan-Meier survival curves tested the unadjusted association of mortality between cemented and uncemented procedures. Multilevel logistic regression, adjusted for covariates, tested the association between cement use and 30-day mortality following arthroplasty. Given the known institutional variation in preference for cemented fixation, an instrumental variable analysis was also performed to minimize the effect of unknown confounders. Adjusted Cox modelling analyzed the association between cement use and mortality at 30 days and one year following surgery.Aims
Methods
It is important to analyze objectively the hammering sound in cup press-fit technique in total hip arthroplasty (THA) in order to better understand the change of the sound during impaction. We hypothesized that a specific characteristic would present in a hammering sound with successful fixation. We designed the study to quantitatively investigate the acoustic characteristics during cementless cup impaction in THA. In 52 THAs performed between November 2018 and April 2022, the acoustic parameters of the hammering sound of 224 impacts of successful press-fit fixation, and 55 impacts of unsuccessful press-fit fixation, were analyzed. The successful fixation was defined if the following two criteria were met: 1) intraoperatively, the stability of the cup was retained after manual application of the torque test; and 2) at one month postoperatively, the cup showed no translation on radiograph. Each hammering sound was converted to sound pressures in 24 frequency bands by fast Fourier transform analysis. Basic patient characteristics were assessed as potential contributors to the hammering sound.Aims
Methods
It is important to analyze objectively the hammering sound in cup press-fit technique in total hip arthroplasty (THA) in order to better understand the change of the sound during impaction. We hypothesized that a specific characteristic would present in a hammering sound with successful fixation. We designed the study to quantitatively investigate the acoustic characteristics during cementless cup impaction in THA. In 52 THAs performed between November 2018 and April 2022, the acoustic parameters of the hammering sound of 224 impacts of successful press-fit fixation, and 55 impacts of unsuccessful press-fit fixation, were analyzed. The successful fixation was defined if the following two criteria were met: 1) intraoperatively, the stability of the cup was retained after manual application of the torque test; and 2) at one month postoperatively, the cup showed no translation on radiograph. Each hammering sound was converted to sound pressures in 24 frequency bands by fast Fourier transform analysis. Basic patient characteristics were assessed as potential contributors to the hammering sound.Aims
Methods
Custom-made partial pelvis replacements (PPRs) are increasingly used in the reconstruction of large acetabular defects and have mainly been designed using a triflange approach, requiring extensive soft-tissue dissection. The monoflange design, where primary intramedullary fixation within the ilium combined with a monoflange for rotational stability, was anticipated to overcome this obstacle. The aim of this study was to evaluate the design with regard to functional outcome, complications, and acetabular reconstruction. Between 2014 and 2023, 79 patients with a mean follow-up of 33 months (SD 22; 9 to 103) were included. Functional outcome was measured using the Harris Hip Score and EuroQol five-dimension questionnaire (EQ-5D). PPR revisions were defined as an endpoint, and subgroups were analyzed to determine risk factors.Aims
Methods
The Peri-Implant and PeriProsthetic Survival AnalysiS (PIPPAS) study aimed to investigate the risk factors for one-year mortality of femoral peri-implant fractures (FPIFs). This prospective, multicentre, observational study involved 440 FPIF patients with a minimum one-year follow-up. Data on demographics, clinical features, fracture characteristics, management, and mortality rates were collected and analyzed using both univariate and multivariate analyses. FPIF patients were elderly (median age 87 years (IQR 81 to 92)), mostly female (82.5%, n = 363), and frail: median clinical frailty scale 6 (IQR 4 to 7), median Pfeiffer 4 (1 to 7), median age-adjusted Charlson Comorbidity Index (CCI) 6 (IQR 5 to 7), and 58.9% (n = 250) were American Society of Anesthesiologists grade III.Aims
Methods
During total knee replacement (TKR), surgeons can choose whether or not to resurface the patella, with advantages and disadvantages of each approach. Recently, the National Institute for Health and Care Excellence (NICE) recommended always resurfacing the patella, rather than never doing so. NICE found insufficient evidence on selective resurfacing (surgeon’s decision based on intraoperative findings and symptoms) to make recommendations. If effective, selective resurfacing could result in optimal individualized patient care. This protocol describes a randomized controlled trial to evaluate the clinical and cost-effectiveness of primary TKR with always patellar resurfacing compared to selective patellar resurfacing. The PAtellar Resurfacing Trial (PART) is a patient- and assessor-blinded multicentre, pragmatic parallel two-arm randomized superiority trial of adults undergoing elective primary TKR for primary osteoarthritis at NHS hospitals in England, with an embedded internal pilot phase (ISRCTN 33276681). Participants will be randomly allocated intraoperatively on a 1:1 basis (stratified by centre and implant type (cruciate-retaining vs cruciate-sacrificing)) to always resurface or selectively resurface the patella, once the surgeon has confirmed sufficient patellar thickness for resurfacing and that constrained implants are not required. The primary analysis will compare the Oxford Knee Score (OKS) one year after surgery. Secondary outcomes include patient-reported outcome measures at three months, six months, and one year (Knee injury and Osteoarthritis Outcome Score, OKS, EuroQol five-dimension five-level questionnaire, patient satisfaction, postoperative complications, need for further surgery, resource use, and costs). Cost-effectiveness will be measured for the lifetime of the patient. Overall, 530 patients will be recruited to obtain 90% power to detect a four-point difference in OKS between the groups one year after surgery, assuming up to 40% resurfacing in the selective group.Aims
Methods
Using data from the Hip Fracture Evaluation with Alternatives of Total Hip Arthroplasty versus Hemiarthroplasty (HEALTH) trial, we sought to determine if a difference in functional outcomes exists between monopolar and bipolar hemiarthroplasty (HA). This study is a secondary analysis of patients aged 50 years or older with a displaced femoral neck fracture who were enrolled in the HEALTH trial and underwent monopolar and bipolar HA. Scores from the Western Ontario and McMaster University Arthritis Index (WOMAC) and 12-Item Short Form Health Survey (SF-12) Physical Component Summary (PCS) and (MCS) were compared between the two HA groups using a propensity score-weighted analysis.Aims
Methods
Periprosthetic fractures (PPFs) following hip arthroplasty are complex injuries. This study evaluates patient demographic characteristics, management, outcomes, and risk factors associated with PPF subtypes over a decade. Using a multicentre collaborative study design, independent of registry data, we identified adults from 29 centres with PPFs around the hip between January 2010 and December 2019. Radiographs were assessed for the Unified Classification System (UCS) grade. Patient and injury characteristics, management, and outcomes were compared between UCS grades. A multinomial logistic regression was performed to estimate relative risk ratios (RRR) of variables on UCS grade.Aims
Methods
Varus-valgus constrained (VVC) devices are typically used in revision settings, often with stems to mitigate the risk of aseptic loosening. However, in at least one system, the VVC insert is compatible with the primary posterior-stabilized (PS) femoral component, which may be an option in complex primary situations. We sought to determine the implant survivorship, radiological and clinical outcomes, and complications when this VVC insert was coupled with a PS femur without stems in complex primary total knee arthroplasties (TKAs). Through our institution’s total joint registry, we identified 113 primary TKAs (103 patients) performed between 2007 and 2017 in which a VVC insert was coupled with a standard cemented PS femur without stems. Mean age was 68 years (SD 10), mean BMI was 32 kg/m2 (SD 7), and 59 patients (50%) were male. Mean follow-up was four years (2 to 10).Aims
Methods
The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems. A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.Aims
Methods
Metal allergy in knee arthroplasty patients is a controversial topic. We aimed to conduct a scoping review to clarify the management of metal allergy in primary and revision total knee arthroplasty (TKA). Studies were identified by searching electronic databases: Cochrane Central Register of Controlled Trials, Ovid MEDLINE, and Embase, from their inception to November 2020, for studies evaluating TKA patients with metal hypersensitivity/allergy. All studies reporting on diagnosing or managing metal hypersensitivity in TKA were included. Data were extracted and summarized based on study design, study population, interventions and outcomes. A practical guide is then formulated based on the available evidence.Aims
Methods
Limb alignment in total knee arthroplasty (TKA) influences periarticular soft-tissue tension, biomechanics through knee flexion, and implant survival. Despite this, there is no uniform consensus on the optimal alignment technique for TKA. Neutral mechanical alignment facilitates knee flexion and symmetrical component wear but forces the limb into an unnatural position that alters native knee kinematics through the arc of knee flexion. Kinematic alignment aims to restore native limb alignment, but the safe ranges with this technique remain uncertain and the effects of this alignment technique on component survivorship remain unknown. Anatomical alignment aims to restore predisease limb alignment and knee geometry, but existing studies using this technique are based on cadaveric specimens or clinical trials with limited follow-up times. Functional alignment aims to restore the native plane and obliquity of the joint by manipulating implant positioning while limiting soft tissue releases, but the results of high-quality studies with long-term outcomes are still awaited. The drawbacks of existing studies on alignment include the use of surgical techniques with limited accuracy and reproducibility of achieving the planned alignment, poor correlation of intraoperative data to long-term functional outcomes and implant survivorship, and a paucity of studies on the safe ranges of limb alignment. Further studies on alignment in TKA should use surgical adjuncts (e.g. robotic technology) to help execute the planned alignment with improved accuracy, include intraoperative assessments of knee biomechanics and periarticular soft-tissue tension, and correlate alignment to long-term functional outcomes and survivorship.
Hip arthroplasty does not always restore normal anatomy. This is due to inaccurate surgery or lack of stem sizes. We evaluated the aptitude of four total hip arthroplasty systems to restore an anatomical and medialized hip rotation centre. Using 3D templating software in 49 CT scans of non-deformed femora, we virtually implanted: 1) small uncemented calcar-guided stems with two offset options (Optimys, Mathys), 2) uncemented straight stems with two offset options (Summit, DePuy Synthes), 3) cemented undersized stems (Exeter philosophy) with three offset options (CPT, ZimmerBiomet), and 4) cemented line-to-line stems (Kerboul philosophy) with proportional offsets (Centris, Mathys). We measured the distance between the templated and the anatomical and 5 mm medialized hip rotation centre.Aims
Methods
This study evaluates risk factors influencing fracture characteristics for postoperative periprosthetic femoral fractures (PFFs) around cemented stems in total hip arthroplasty. Data were collected for PFF patients admitted to eight UK centres between 25 May 2006 and 1 March 2020. Radiographs were assessed for Unified Classification System (UCS) grade and AO/OTA type. Statistical comparisons investigated relationships by age, gender, and stem fixation philosophy (polished taper-slip (PTS) vs composite beam (CB)). The effect of multiple variables was estimated using multinomial logistic regression to estimate odds ratios (ORs) with 95% confidence intervals (CIs). Surgical treatment (revision vs fixation) was compared by UCS grade and AO/OTA type.Aims
Methods
Complex total hip arthroplasty (THA) with subtrochanteric shortening osteotomy is necessary in conditions other than developmental dysplasia of the hip (DDH) and septic arthritis sequelae with significant proximal femur migration. Our aim was to evaluate the hip centre restoration with THAs in these hips. In all, 27 THAs in 25 patients requiring THA with femoral shortening between 2012 and 2019 were assessed. Bilateral shortening was required in two patients. Subtrochanteric shortening was required in 14 out of 27 hips (51.9%) with aetiology other than DDH or septic arthritis. Vertical centre of rotation (VCOR), horizontal centre of rotation, offset, and functional outcome was calculated. The mean followup was 24.4 months (5 to 92 months).Aims
Methods
Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated. This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (Objectives
Methods
The medially spherical GMK Sphere (Medacta International AG, Castel San Pietro, Switzerland) total knee arthroplasty (TKA) was previously shown to accommodate lateral rollback while pivoting around a stable medial compartment, aiming to replicate native knee kinematics in which some coronal laxity, especially laterally, is also present. We assess coronal plane kinematics of the GMK Sphere and explore the occurrence and pattern of articular separation during static and dynamic activities. Using pulsed fluoroscopy and image matching, the coronal kinematics and articular surface separation of 16 well-functioning TKAs were studied during weight-bearing and non-weight-bearing, static, and dynamic activities. The closest distances between the modelled articular surfaces were examined with respect to knee position, and proportions of joint poses exhibiting separation were computed.Objectives
Methods
Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient’s surrounding tissues. A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient’s tissues.Objectives
Methods
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods
Malrotation of the femoral component can result in post-operative complications in total knee arthroplasty (TKA), including patellar maltracking. Therefore, we used computational simulation to investigate the influence of femoral malrotation on contact stresses on the polyethylene (PE) insert and on the patellar button as well as on the forces on the collateral ligaments. Validated finite element (FE) models, for internal and external malrotations from 0° to 10° with regard to the neutral position, were developed to evaluate the effect of malrotation on the femoral component in TKA. Femoral malrotation in TKA on the knee joint was simulated in walking stance-phase gait and squat loading conditions.Objectives
Materials and Methods
Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep. We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem.Objectives
Methods
Infection of implants is a major problem in elective and trauma surgery. Heating is an effective way to reduce the bacterial load in food preparation, and studies on hyperthermia treatment for cancer have shown that it is possible to heat metal objects with pulsed electromagnetic fields selectively (PEMF), also known as induction heating. We therefore set out to answer the following research question: is non-contact induction heating of metallic implants effective in reducing bacterial load Titanium alloy cylinders (Ti6Al4V) were exposed to PEMF from an induction heater with maximum 2000 watts at 27 kHz after being contaminated with five different types of micro-organisms: Objectives
Methods
Since redesign of the Oxford phase III mobile-bearing unicompartmental
knee arthroplasty (UKA) femoral component to a twin-peg design,
there has not been a direct comparison to total knee arthroplasty
(TKA). Thus, we explored differences between the two cohorts. A total of 168 patients (201 knees) underwent medial UKA with
the Oxford Partial Knee Twin-Peg. These patients were compared with
a randomly selected group of 177 patients (189 knees) with primary
Vanguard TKA. Patient demographics, Knee Society (KS) scores and
range of movement (ROM) were compared between the two cohorts. Additionally,
revision, re-operation and manipulation under anaesthesia rates
were analysed.Aims
Patients and Methods
Because there have been no standard methods to determine pre-operatively
the thickness of resection of the proximal tibia in unicompartmental
knee arthroplasty (UKA), information about the relationship between
the change of limb alignment and the joint line elevation would
be useful for pre-operative planning. The purpose of this study
was to clarify the correlation between the change of limb alignment
and the change of joint line height at the medial compartment after
UKA. A consecutive series of 42 medial UKAs was reviewed retrospectively.
These patients were assessed radiographically both pre- and post-operatively
with standing anteroposterior radiographs. The thickness of bone
resection at the proximal tibia and the distal femur was measured
radiographically. The relationship between the change of femorotibial
angle (δFTA) and the change of joint line height, was analysed.Objectives
Methods
Throughout the 20th Century, it has been postulated that the knee moves on the basis of a four-bar link mechanism composed of the cruciate ligaments, the femur and the tibia. As a consequence, the femur has been thought to roll back with flexion, and total knee arthroplasty (TKA) prostheses have been designed on this basis. Recent work, however, has proposed that at a position of between 0° and 120° the medial femoral condyle does not move anteroposteriorly whereas the lateral femoral condyle tends, but is not obliged, to roll back – a combination of movements which equates to tibial internal/ femoral external rotation with flexion. The aim of this paper was to assess if the articular geometry of the GMK Sphere TKA could recreate the natural knee movements The pattern of knee movement was studied in 15 patients (six male: nine female; one male with bilateral TKAs) with 16 GMK Sphere implants, at a mean age of 66 years (53 to 76) with a mean BMI of 30 kg/m2 (20 to 35). The motions of all 16 knees were observed using pulsed fluoroscopy during a number of weight-bearing and non-weight-bearing static and dynamic activities.Objectives
Methods
This study tests the biomechanical properties of adjacent locked
plate constructs in a femur model using Sawbones. Previous studies
have described biomechanical behaviour related to inter-device distances.
We hypothesise that a smaller lateral inter-plate distance will
result in a biomechanically stronger construct, and that addition
of an anterior plate will increase the overall strength of the construct. Sawbones were plated laterally with two large-fragment locking
compression plates with inter-plate distances of 10 mm or 1 mm.
Small-fragment locking compression plates of 7-hole, 9-hole, and
11-hole sizes were placed anteriorly to span the inter-plate distance.
Four-point bend loading was applied, and the moment required to
displace the constructs by 10 mm was recorded.Objectives
Methods
Total knee arthroplasty (TKA) is an established
and successful procedure. However, the design of prostheses continues
to be modified in an attempt to optimise the functional outcome
of the patient. The aim of this study was to determine if patient outcome after
TKA was influenced by the design of the prosthesis used. A total of 212 patients (mean age 69; 43 to 92; 131 female (62%),
81 male (32%)) were enrolled in a single centre double-blind trial
and randomised to receive either a Kinemax (group 1) or a Triathlon
(group 2) TKA. Patients were assessed pre-operatively, at six weeks, six months,
one year and three years after surgery. The outcome assessments
used were the Oxford Knee Score; range of movement; pain numerical
rating scales; lower limb power output; timed functional assessment
battery and a satisfaction survey. Data were assessed incorporating
change over all assessment time points, using repeated measures
analysis of variance longitudinal mixed models. Implant group 2
showed a significantly greater range of movement (p = 0.009), greater
lower limb power output (p = 0.026) and reduced report of ‘worst
daily pain’ (p = 0.003) over the three years of follow-up. Differences
in Oxford Knee Score (p = 0.09), report of ‘average daily pain’
(p = 0.57) and timed functional performance tasks (p = 0.23) did
not reach statistical significance. Satisfaction with outcome was
significantly better in group 2 (p = 0.001). These results suggest that patient outcome after TKA can be influenced
by the prosthesis used. Cite this article:
Fractures of the proximal femur are one of the
greatest challenges facing the medical community, constituting a
heavy socioeconomic burden worldwide. Controversy exists regarding
the optimal treatment for independent patients with displaced intracapsular fractures
of the proximal femur. The recognised alternatives are hemiarthroplasty
and total hip replacement. At present there is no established standard
of care, with both types of arthroplasty being used in many centres.
The principal advantages of total hip replacement are a functional
benefit over hemiarthroplasty and a reduced risk of revision surgery.
The principal criticism is the increased risk of dislocation. We
believe that an alternative acetabular component may reduce the
risk of dislocation but still provide the functional benefit of
total hip replacement in these patients. We therefore propose to
investigate the dislocation risk of a dual-mobility acetabular component
compared with standard polyethylene component in total hip replacement
for independent patients with displaced intracapsular fractures
of the proximal femur within the framework of the larger WHiTE (Warwick
Hip Trauma Evaluation) Comprehensive Cohort Study. Cite this article:
To study the vascularity and bone metabolism of the femoral head/neck
following hip resurfacing arthroplasty, and to use these results
to compare the posterior and the trochanteric-flip approaches. In our previous work, we reported changes to intra-operative
blood flow during hip resurfacing arthroplasty comparing two surgical
approaches. In this study, we report the vascularity and the metabolic
bone function in the proximal femur in these same patients at one
year after the surgery. Vascularity and bone function was assessed
using scintigraphic techniques. Of the 13 patients who agreed to
take part, eight had their arthroplasty through a posterior approach
and five through a trochanteric-flip approach.Objectives
Methods
Orthopaedic surgeons use stems in revision knee surgery to obtain
stability when metaphyseal bone is missing. No consensus exists
regarding stem size or method of fixation. This A custom test rig using differential variable reluctance transducers
(DVRTs) was developed to record all translational and rotational
motions at the bone–implant interface. Composite femurs were used.
These were secured to permit variation in flexion angle from 0°
to 90°. Cyclic loads were applied through a tibial component based
on three peaks corresponding to 0°, 10° and 20° flexion from a normal
walking cycle. Three different femoral components were investigated
in this study for cementless and cemented interface conditions.Objectives
Methods