header advert
Results 1 - 20 of 36
Results per page:
Bone & Joint Research
Vol. 11, Issue 10 | Pages 690 - 699
4 Oct 2022
Lenguerrand E Whitehouse MR Kunutsor SK Beswick AD Baker RP Rolfson O Reed MR Blom AW

Aims. We compared the risks of re-revision and mortality between two-stage revision surgery and single-stage revision surgery among patients with infected primary knee arthroplasty. Methods. Patients with a periprosthetic joint infection (PJI) of their primary knee arthroplasty, initially revised with a single-stage or a two-stage procedure in England and Wales between 2003 and 2014, were identified from the National Joint Registry. We used Poisson regression with restricted cubic splines to compute hazard ratios (HR) at different postoperative periods. The total number of revisions and re-revisions undergone by patients was compared between the two strategies. Results. A total of 489 primary knee arthroplasties were revised with single-stage procedure (1,390 person-years) and 2,377 with two-stage procedure (8,349 person-years). The adjusted incidence rates of all-cause re-revision and for infection were comparable between these strategies (HR overall five years, 1.15 (95% confidence interval (CI) 0.87 to 1.52), p = 0.308; HR overall five years, 0.99 (95% CI 0.70 to 1.39), p = 0.949, respectively). Patients initially managed with single-stage revision received fewer revision procedures overall than after two-stage revision (1.2 vs 2.2, p < 0.001). Mortality was lower for single-stage revision between six and 18 months postoperative (HR at six months, 0.51 (95% CI 0.25 to 1.00), p = 0.049 HR at 18 months, 0.33 (95% CI 0.12 to 0.99), p = 0.048) and comparable at other timepoints. Conclusion. The risk of re-revision was similar between single- and two-stage revision for infected primary knee arthroplasty. Single-stage group required fewer revisions overall, with lower or comparable mortality at specific postoperative periods. The single-stage revision is a safe and effective strategy to treat infected knee arthroplasties. There is potential for increased use to reduce the burden of knee PJI for patients, and for the healthcare system. Cite this article: Bone Joint Res 2022;11(10):690–699


Bone & Joint Research
Vol. 13, Issue 8 | Pages 392 - 400
5 Aug 2024
Barakat A Evans J Gibbons C Singh HP

Aims. The Oxford Shoulder Score (OSS) is a 12-item measure commonly used for the assessment of shoulder surgeries. This study explores whether computerized adaptive testing (CAT) provides a shortened, individually tailored questionnaire while maintaining test accuracy. Methods. A total of 16,238 preoperative OSS were available in the National Joint Registry (NJR) for England, Wales, Northern Ireland, the Isle of Man, and the States of Guernsey dataset (April 2012 to April 2022). Prior to CAT, the foundational item response theory (IRT) assumptions of unidimensionality, monotonicity, and local independence were established. CAT compared sequential item selection with stopping criteria set at standard error (SE) < 0.32 and SE < 0.45 (equivalent to reliability coefficients of 0.90 and 0.80) to full-length patient-reported outcome measure (PROM) precision. Results. Confirmatory factor analysis (CFA) for unidimensionality exhibited satisfactory fit with root mean square standardized residual (RSMSR) of 0.06 (cut-off ≤ 0.08) but not with comparative fit index (CFI) of 0.85 or Tucker-Lewis index (TLI) of 0.82 (cut-off > 0.90). Monotonicity, measured by H value, yielded 0.482, signifying good monotonic trends. Local independence was generally met, with Yen’s Q3 statistic > 0.2 for most items. The median item count for completing the CAT simulation with a SE of 0.32 was 3 (IQR 3 to 12), while for a SE of 0.45 it was 2 (IQR 2 to 6). This constituted only 25% and 16%, respectively, when compared to the 12-item full-length questionnaire. Conclusion. Calibrating IRT for the OSS has resulted in the development of an efficient and shortened CAT while maintaining accuracy and reliability. Through the reduction of redundant items and implementation of a standardized measurement scale, our study highlights a promising approach to alleviate time burden and potentially enhance compliance with these widely used outcome measures. Cite this article: Bone Joint Res 2024;13(8):392–400


Bone & Joint Research
Vol. 12, Issue 5 | Pages 321 - 330
9 May 2023
Lenguerrand E Whitehouse MR Beswick AD Kunutsor SK Webb JCJ Mehendale S Porter M Blom AW

Aims. We compared the risks of re-revision and mortality between two-stage and single-stage revision surgeries among patients with infected primary hip arthroplasty. Methods. Patients with a periprosthetic joint infection (PJI) of their primary arthroplasty revised with single-stage or two-stage procedure in England and Wales between 2003 and 2014 were identified from the National Joint Registry. We used Poisson regression with restricted cubic splines to compute hazard ratios (HRs) at different postoperative periods. The total number of revisions and re-revisions undergone by patients was compared between the two strategies. Results. In total, 535 primary hip arthroplasties were revised with single-stage procedure (1,525 person-years) and 1,605 with two-stage procedure (5,885 person-years). All-cause re-revision was higher following single-stage revision, especially in the first three months (HR at 3 months = 1.98 (95% confidence interval (CI) 1.14 to 3.43), p = 0.009). The risks were comparable thereafter. Re-revision for PJI was higher in the first three postoperative months for single-stage revision and waned with time (HR at 3 months = 1.81 (95% CI 1.22 to 2.68), p = 0.003; HR at 6 months = 1.25 (95% CI 0.71 to 2.21), p = 0.441; HR at 12 months = 0.94 (95% CI 0.54 to 1.63), p = 0.819). Patients initially managed with a single-stage revision received fewer revision operations (mean 1.3 (SD 0.7) vs 2.2 (SD 0.6), p < 0.001). Mortality rates were comparable between these two procedures (29/10,000 person-years vs 33/10,000). Conclusion. The risk of unplanned re-revision was lower following two-stage revision, but only in the early postoperative period. The lower overall number of revision procedures associated with a single-stage revision strategy and the equivalent mortality rates to two-stage revision are reassuring. With appropriate counselling, single-stage revision is a viable option for the treatment of hip PJI. Cite this article: Bone Joint Res 2023;12(5):321–330


Bone & Joint Research
Vol. 12, Issue 4 | Pages 256 - 258
3 Apr 2023
Farrow L Evans J

Cite this article: Bone Joint Res 2023;12(4):256–258.


Aims. To investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs). Methods. In this registry study, data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man (NJR) were linked with manufacturing data supplied by manufacturers. The primary endpoint was revision of any component. Cox proportional hazard regression was a primary analytic approach adjusting for competing risk of death, patient characteristics, head composition, and stem fixation. Results. A total of 290,770 primary THAs were successfully linked with manufacturing characteristics. Overall 4,708 revisions were analyzed, 1,260 of which were due to aseptic loosening. Total radiation dose was identified as a risk factor and included in the Cox model. For statistical modelling of aseptic loosening, THAs were grouped into three categories: G1 (no radiation); G2 ( > 0 to < 5 Mrad); and G3 ( ≥ 5 Mrad). G1 had the worst survivorship. The Cox regression hazard ratio for revision due to aseptic loosening for G2 was 0.7 (95% confidence interval (CI) 0.58 to 0.83), and for G3 0.4 (95% CI 0.30 to 0.53). Male sex and uncemented stem fixation were associated with higher risk of revision and ceramic heads with lower risk. Conclusion. Polyethylene irradiation was associated with reduced risk of revision for aseptic loosening. Radiation doses of ≥ 5 Mrad were associated with a further reduction in risk. Cite this article: Bone Joint Res 2020;9(9):563–571


Bone & Joint Research
Vol. 6, Issue 6 | Pages 391 - 398
1 Jun 2017
Lenguerrand E Whitehouse MR Beswick AD Jones SA Porter ML Blom* AW

Objectives. We used the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man (NJR) to investigate the risk of revision due to prosthetic joint infection (PJI) for patients undergoing primary and revision hip arthroplasty, the changes in risk over time, and the overall burden created by PJI. Methods. We analysed revision total hip arthroplasties (THAs) performed due to a diagnosis of PJI and the linked index procedures recorded in the NJR between 2003 and 2014. The cohort analysed consisted of 623 253 index primary hip arthroplasties, 63 222 index revision hip arthroplasties and 7585 revision THAs performed due to a diagnosis of PJI. The prevalence, cumulative incidence functions and the burden of PJI (total procedures) were calculated. Overall linear trends were investigated with log-linear regression. Results. We demonstrated a prevalence of revision THA due to prosthetic joint infection of 0.4/100 procedures following primary and 1.6/100 procedures following revision hip arthroplasty. The prevalence of revision due to PJI in the three months following primary hip arthroplasty has risen 2.3-fold (95% confidence interval (CI) 1.3 to 4.1) between 2005 and 2013, and 3.0-fold (95% CI 1.1 to 8.5) following revision hip arthroplasty. Over 1000 procedures are performed annually as a consequence of hip PJI, an increase of 2.6-fold between 2005 and 2013. Conclusions. Although the risk of revision due to PJI following hip arthroplasty is low, it is rising and, coupled with the established and further predicted increased incidence of both primary and revision hip arthroplasty, this represents a growing and substantial treatment burden. Cite this article: E. Lenguerrand, M. R. Whitehouse, A. D. Beswick, S. A. Jones, M. L. Porter, A. W. Blom. Revision for prosthetic joint infection following hip arthroplasty: Evidence from the National Joint Registry. Bone Joint Res 2017;6:391–398. DOI: 10.1302/2046-3758.66.BJR-2017-0003.R1


Bone & Joint Research
Vol. 5, Issue 6 | Pages 206 - 214
1 Jun 2016
Malak TT Broomfield JAJ Palmer AJR Hopewell S Carr A Brown C Prieto-Alhambra D Glyn-Jones S

Objectives. High failure rates of metal-on-metal hip arthroplasty implants have highlighted the need for more careful introduction and monitoring of new implants and for the evaluation of the safety of medical devices. The National Joint Registry and other regulatory services are unable to detect failing implants at an early enough stage. We aimed to identify validated surrogate markers of long-term outcome in patients undergoing primary total hip arthroplasty (THA). Methods. We conducted a systematic review of studies evaluating surrogate markers for predicting long-term outcome in primary THA. Long-term outcome was defined as revision rate of an implant at ten years according to National Institute of Health and Care Excellence guidelines. We conducted a search of Medline and Embase (OVID) databases. Separate search strategies were devised for the Cochrane database and Google Scholar. Each search was performed to include articles from the date of their inception to June 8, 2015. Results. Our search strategy identified 1082 studies of which 115 studies were included for full article review. Following review, 17 articles were found that investigated surrogate markers of long-term outcome. These included one systematic review, one randomised control trial (RCT), one case control study and 13 case series. Validated surrogate markers included Radiostereometric Analysis (RSA) and Einzel-Bild-Röntgen-Analyse (EBRA), each measuring implant migration and wear. We identified five RSA studies (one systematic review and four case series) and four EBRA studies (one RCT and three case series). Patient Reported Outcome Measures (PROMs) at six months have been investigated but have not been validated against long-term outcomes. Conclusions. This systematic review identified two validated surrogate markers of long-term primary THA outcome: RSA and EBRA, each measuring implant migration and wear. We recommend the consideration of RSA in the pre-market testing of new implants. EBRA can be used to investigate acetabular wear but not femoral migration. Further studies are needed to validate the use of PROMs for post-market surveillance. Cite this article: T. T. Malak, J. A. J. Broomfield, A. J. R. Palmer, S. Hopewell, A. Carr, C. Brown, D. Prieto-Alhambra, S. Glyn-Jones. Surrogate markers of long-term outcome in primary total hip arthroplasty: A systematic review. Bone Joint Res 2016;5:206–214. DOI: 10.1302/2046-3758.56.2000568


Bone & Joint Research
Vol. 3, Issue 4 | Pages 108 - 116
1 Apr 2014
Cheng K Giebaly D Campbell A Rumley A Lowe G

Objective. Mortality rates reported by the National Joint Registry for England and Wales (NJR) were higher following cemented total knee replacement (TKR) compared with uncemented procedures. The aim of this study is to examine and compare the effects of cemented and uncemented TKR on the activation of selected markers of inflammation, endothelium, and coagulation, and on the activation of selected cytokines involved in the various aspects of the systemic response following surgery. Methods. This was a single centre, prospective, case-control study. Following enrolment, blood samples were taken pre-operatively, and further samples were collected at day one and day seven post-operatively. One patient in the cemented group developed a deep-vein thrombosis confirmed on ultrasonography and was excluded, leaving 19 patients in this cohort (mean age 67.4, (. sd. 10.62)), and one patient in the uncemented group developed a post-operative wound infection and was excluded, leaving 19 patients (mean age 66.5, (. sd. 7.82)). Results. Both groups had a similar response with regards to the levels of C-reactive protein (CRP), interleukin 6 (IL-6) and tumour necrosis factor-alpha (TNFα). CD40 levels rose significantly on the cemented group over day one to day seven compared with that of the uncemented group, which occurred over the first 24 hours. The CD14/42a levels demonstrated a statistically significant increase in the cemented group (p < 0.001 first 24 hours and p = 0.02 between days one and seven). . Conclusions. The uncemented and cemented groups demonstrated significant changes in the various parameters measured at various time points but apart from CD14/42a levels, there was no significant difference in the serum markers of inflammation, coagulation and endothelial dysfunction following cemented TKR. Cite this article: Bone Joint Res 2014;3:108–16


Bone & Joint Research
Vol. 13, Issue 5 | Pages 201 - 213
1 May 2024
Hamoodi Z Gehringer CK Bull LM Hughes T Kearsley-Fleet L Sergeant JC Watts AC

Aims

The aims of this study were to identify and evaluate the current literature examining the prognostic factors which are associated with failure of total elbow arthroplasty (TEA).

Methods

Electronic literature searches were conducted using MEDLINE, Embase, PubMed, and Cochrane. All studies reporting prognostic estimates for factors associated with the revision of a primary TEA were included. The risk of bias was assessed using the Quality In Prognosis Studies (QUIPS) tool, and the quality of evidence was assessed using the modified Grading of Recommendations, Assessment, Development, and Evaluations (GRADE) framework. Due to low quality of the evidence and the heterogeneous nature of the studies, a narrative synthesis was used.


Bone & Joint Research
Vol. 12, Issue 5 | Pages 331 - 338
16 May 2023
Szymski D Walter N Krull P Melsheimer O Grimberg A Alt V Steinbrueck A Rupp M

Aims

The aim of this investigation was to compare risk of infection in both cemented and uncemented hemiarthroplasty (HA) as well as in total hip arthroplasty (THA) following femoral neck fracture.

Methods

Data collection was performed using the German Arthroplasty Registry (EPRD). In HA and THA following femoral neck fracture, fixation method was divided into cemented and uncemented prostheses and paired according to age, sex, BMI, and the Elixhauser Comorbidity Index using Mahalanobis distance matching.


Bone & Joint Research
Vol. 11, Issue 5 | Pages 270 - 277
6 May 2022
Takegami Y Seki T Osawa Y Imagama S

Aims

Periprosthetic hip fractures (PPFs) after total hip arthroplasty are difficult to treat. Therefore, it is important to identify modifiable risk factors such as stem selection to reduce the occurrence of PPFs. This study aimed to clarify differences in fracture torque, surface strain, and fracture type analysis between three different types of cemented stems.

Methods

We conducted biomechanical testing of bone analogues using six cemented stems of three different types: collarless polished tapered (CPT) stem, Versys Advocate (Versys) stem, and Charnley-Marcel-Kerboull (CMK) stem. Experienced surgeons implanted each of these types of stems into six bone analogues, and the analogues were compressed and internally rotated until failure. Torque to fracture and fracture type were recorded. We also measured surface strain distribution using triaxial rosettes.


Bone & Joint Research
Vol. 10, Issue 9 | Pages 591 - 593
7 Sep 2021
Thompson JW Simpson AHRW Haddad FS


Bone & Joint Research
Vol. 10, Issue 6 | Pages 340 - 347
1 Jun 2021
Jenkinson MRJ Meek RMD Tate R MacMillan S Grant MH Currie S

Elevated levels of circulating cobalt ions have been linked with a wide range of systemic complications including neurological, endocrine, and cardiovascular symptoms. Case reports of patients with elevated blood cobalt ions have described significant cardiovascular complications including cardiomyopathy. However, correlation between the actual level of circulating cobalt and extent of cardiovascular injury has not previously been performed. This review examines evidence from the literature for a link between elevated blood cobalt levels secondary to metal-on-metal (MoM) hip arthroplasties and cardiomyopathy. Correlation between low, moderate, and high blood cobalt with cardiovascular complications has been considered. Elevated blood cobalt at levels over 250 µg/l have been shown to be a risk factor for developing systemic complications and published case reports document cardiomyopathy, cardiac transplantation, and death in patients with severely elevated blood cobalt ions. However, it is not clear that there is a hard cut-off value and cardiac dysfunction may occur at lower levels. Clinical and laboratory research has found conflicting evidence of cobalt-induced cardiomyopathy in patients with MoM hips. Further work needs to be done to clarify the link between severely elevated blood cobalt ions and cardiomyopathy.

Cite this article: Bone Joint Res 2021;10(6):340–347.


Bone & Joint Research
Vol. 10, Issue 6 | Pages 348 - 350
1 Jun 2021
Skinner JA Sabah SA Hart AJ


Bone & Joint Research
Vol. 8, Issue 6 | Pages 228 - 231
1 Jun 2019
Kayani B Haddad FS


Bone & Joint Research
Vol. 8, Issue 11 | Pages 570 - 572
1 Nov 2019
Ahmed SS Haddad FS


Bone & Joint Research
Vol. 8, Issue 6 | Pages 246 - 252
1 Jun 2019
Liddle A Webb M Clement N Green S Liddle J German M Holland J

Objectives

Previous studies have evidenced cement-in-cement techniques as reliable in revision arthroplasty. Commonly, the original cement mantle is reshaped, aiding accurate placement of the new stem. Ultrasonic devices selectively remove cement, preserve host bone, and have lower cortical perforation rates than other techniques. As far as the authors are aware, the impact of ultrasonic devices on final cement-in-cement bonds has not been investigated. This study assessed the impact of cement removal using the Orthosonics System for Cemented Arthroplasty Revision (OSCAR; Orthosonics) on final cement-in-cement bonds.

Methods

A total of 24 specimens were manufactured by pouring cement (Simplex P Bone Cement; Stryker) into stainless steel moulds, with a central rod polished to Stryker Exeter V40 specifications. After cement curing, the rods were removed and eight specimens were allocated to each of three internal surface preparation groups: 1) burr; 2) OSCAR; and 3) no treatment. Internal holes were recemented, and each specimen was cut into 5 mm discs. Shear testing of discs was completed by a technician blinded to the original grouping, recording ultimate shear strengths. Scanning electron microscopy (SEM) was completed, inspecting surfaces of shear-tested specimens.


Bone & Joint Research
Vol. 7, Issue 12 | Pages 636 - 638
1 Dec 2018
Roussot MA Haddad FS


Bone & Joint Research
Vol. 7, Issue 11 | Pages 595 - 600
1 Nov 2018
Bergiers S Hothi HS Henckel J Eskelinen A Skinner J Hart A

Objectives

Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event.

Methods

A total of 92 retrieved MoM Pinnacle hips were analyzed, of which 45 were implanted before 2007, and 47 from 2007 onwards. The ‘pre-2007’ group contained 45 implants retrieved from 21 male and 24 female patients, with a median age of 61.3 years (interquartile range (IQR) 57.1 to 65.5); the ‘2007 onwards’ group contained 47 implants retrieved from 19 male and 28 female patients, with a median age of 61.8 years (IQR 58.5 to 67.8). The volume of material lost from their bearing and taper surfaces was measured using coordinate and roundness measuring machines. These outcomes were then compared statistically using linear regression models, adjusting for potentially confounding factors.


Bone & Joint Research
Vol. 7, Issue 8 | Pages 508 - 510
1 Aug 2018
Horriat S Haddad FS