Aims. The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years. Methods. We reviewed the results of 165 THAs using the Omnifit HA system in 138 patients, performed between August 1993 and December 1999. The mean age of the patients at the time of surgery was 46 years (20 to 77). Avascular necrosis was the most common indication for THA, followed by ankylosing spondylitis and primary osteoarthritis. The mean follow-up was 22 years (20 to 31). At 20 and 25 years, 113 THAs in 91 patients and 63 THAs in 55 patients were available for review, respectively, while others died or were lost to follow-up.
Aims. Hip fractures are some of the most common fractures encountered in orthopaedic practice. We aimed to identify whether perioperative hypotension is a predictor of 30-day mortality, and to stratify patient groups that would benefit from closer monitoring and early intervention. While there is literature on intraoperative blood pressure, there are limited studies examining pre- and postoperative blood pressure. Methods. We conducted a prospective observational cohort study over a one-year period from December 2021 to December 2022. Patient demographic details, biochemical results, and haemodynamic observations were taken from electronic medical records. Statistical analysis was conducted with the Cox proportional hazards model, and the effects of independent variables estimated with the Wald statistic.
Aims. The aim of this study was to investigate the association between fracture displacement and survivorship of the native hip joint without conversion to a total hip arthroplasty (THA), and to determine predictors for conversion to THA in patients treated nonoperatively for acetabular fractures. Methods. A multicentre cross-sectional study was performed in 170 patients who were treated nonoperatively for an acetabular fracture in three level 1 trauma centres. Using the post-injury diagnostic CT scan, the maximum gap and step-off values in the weightbearing dome were digitally measured by two trauma surgeons. Native hip survival was reported using
We compared decompression alone to decompression with fusion surgery for lumbar spinal stenosis, with or without degenerative spondylolisthesis (DS). The aim was to evaluate if five-year outcomes differed between the groups. The two-year results from the same trial revealed no differences. The Swedish Spinal Stenosis Study was a multicentre randomized controlled trial with recruitment from September 2006 to February 2012. A total of 247 patients with one- or two-level central lumbar spinal stenosis, stratified by the presence of DS, were randomized to decompression alone or decompression with fusion. The five-year Oswestry Disability Index (ODI) was the primary outcome. Secondary outcomes were the EuroQol five-dimension questionnaire (EQ-5D), visual analogue scales for back and leg pain, and patient-reported satisfaction, decreased pain, and increased walking distance. The reoperation rate was recorded.Aims
Methods
We assessed the value of the Clinical Frailty Scale (CFS) in the prediction of adverse outcome after hip fracture. Of 1,577 consecutive patients aged > 65 years with a fragility hip fracture admitted to one institution, for whom there were complete data, 1,255 (72%) were studied. Clinicians assigned CFS scores on admission. Audit personnel routinely prospectively completed the Standardised Audit of Hip Fracture in Europe form, including the following outcomes: 30-day survival; in-hospital complications; length of acute hospital stay; and new institutionalization. The relationship between the CFS scores and outcomes was examined graphically and the visual interpretations were tested statistically. The predictive values of the CFS and Nottingham Hip Fracture Score (NHFS) to predict 30-day mortality were compared using receiver operating characteristic area under the curve (AUC) analysis.Aims
Methods
Isolated acetabular liner exchange with a highly crosslinked polyethylene (HXLPE) component is an option to address polyethylene wear and osteolysis following total hip arthroplasty (THA) in the presence of a well-fixed acetabular shell. The liner can be fixed either with the original locking mechanism or by being cemented within the acetabular component. Whether the method used for fixation of the HXLPE liner has any bearing on the long-term outcomes is still unclear. Data were retrieved for all patients who underwent isolated acetabular component liner exchange surgery with a HXLPE component in our institute between August 2000 and January 2015. Patients were classified according to the fixation method used (original locking mechanism (n = 36) or cemented (n = 50)). Survival and revision rates were compared. A total of 86 revisions were performed and the mean duration of follow-up was 13 years.Aims
Methods
Despite higher rates of revision after total hip arthroplasty (THA) being reported for uncemented stems in patients aged > 75 years, they are frequently used in this age group. Increased mortality after cemented fixation is often used as a justification, but recent data do not confirm this association. The aim of this study was to investigate the influence of the design of the stem and the type of fixation on the rate of revision and immediate postoperative mortality, focusing on the age and sex of the patients. A total of 333,144 patients with primary osteoarthritis (OA) of the hip who underwent elective THA between November 2012 and September 2022, using uncemented acetabular components without reconstruction shells, from the German arthroplasty registry were included in the study. The revision rates three years postoperatively for four types of stem (uncemented, uncemented with collar, uncemented short, and cemented) were compared within four age groups: < 60 years (Young), between 61 and 70 years (Mid-I), between 71 and 80 years (Mid-II), and aged > 80 years (Old). A noninferiority analysis was performed on the most frequently used designs of stem.Aims
Methods
The aim of this study was to estimate the clinical and economic burden of dislocation following primary total hip arthroplasty (THA) in England. This retrospective evaluation used data from the UK Clinical Practice Research Datalink database. Patients were eligible if they underwent a primary THA (index date) and had medical records available 90 days pre-index and 180 days post-index. Bilateral THAs were excluded. Healthcare costs and resource use were evaluated over two years. Changes (pre- vs post-THA) in generic quality of life (QoL) and joint-specific disability were evaluated. Propensity score matching controlled for baseline differences between patients with and without THA dislocation.Aims
Methods
Excision of chronic osteomyelitic bone creates a dead space which must be managed to avoid early recurrence of infection. Systemic antibiotics cannot penetrate this space in high concentrations, so local treatment has become an attractive adjunct to surgery. The aim of this study was to present the mid- to long-term results of local treatment with gentamicin in a bioabsorbable ceramic carrier. A prospective series of 100 patients with Cierny-Mader Types III and IV chronic ostemyelitis, affecting 105 bones, were treated with a single-stage procedure including debridement, deep tissue sampling, local and systemic antibiotics, stabilization, and immediate skin closure. Chronic osteomyelitis was confirmed using strict diagnostic criteria. The mean follow-up was 6.05 years (4.2 to 8.4).Aims
Methods
The aim of this study was to inform the epidemiology and treatment of slipped capital femoral epiphysis (SCFE). This was an anonymized comprehensive cohort study, with a nested consented cohort, following the the Idea, Development, Exploration, Assessment, Long-term study (IDEAL) framework. A total of 143 of 144 hospitals treating SCFE in Great Britain participated over an 18-month period. Patients were cross-checked against national administrative data and potential missing patients were identified. Clinician-reported outcomes were collected until two years. Patient-reported outcome measures (PROMs) were collected for a subset of participants.Aims
Methods
This study aimed to identify risk factors (patient, healthcare system, and socioeconomic) for mortality after hip fractures and estimate their relative importance. Further, we aimed to elucidate mortality and survival patterns following fractures and the duration of excess mortality. Data on 37,394 hip fractures in the Norwegian Hip Fracture Register from January 2014 to December 2018 were linked to data from the Norwegian Patient Registry, Statistics Norway, and characteristics of acute care hospitals. Cox regression analysis was performed to estimate risk factors associated with mortality. The Wald statistic was used to estimate and illustrate relative importance of risk factors, which were categorized in modifiable (healthcare-related) and non-modifiable (patient-related and socioeconomic). We calculated standardized mortality ratios (SMRs) comparing deaths among hip fracture patients to expected deaths in a standardized reference population.Aims
Methods
Unicompartmental knee arthroplasty (UKA) is a bone-preserving treatment option for osteoarthritis localized to a single compartment in the knee. The success of the procedure is sensitive to patient selection and alignment errors. Robotic arm-assisted UKA provides technological assistance to intraoperative bony resection accuracy, which is thought to improve ligament balancing. This paper presents the five-year outcomes of a comparison between manual and robotically assisted UKAs. The trial design was a prospective, randomized, parallel, single-centre study comparing surgical alignment in patients undergoing UKA for the treatment of medial compartment osteoarthritis (ISRCTN77119437). Participants underwent surgery using either robotic arm-assisted surgery or conventional manual instrumentation. The primary outcome measure (surgical accuracy) has previously been reported, and, along with secondary outcomes, were collected at one-, two-, and five-year timepoints. Analysis of five-year results and longitudinal analysis for all timepoints was performed to compare the two groups.Aims
Methods
The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression.Aims
Methods
Various surgical techniques have been described for total hip arthroplasty (THA) in patients with Crowe type III dislocated hips, who have a large acetabular bone defect. The aim of this study was to evaluate the long-term clinical results of patients in whom anatomical reconstruction of the acetabulum was performed using a cemented acetabular component and autologous bone graft from the femoral neck. A total of 22 patients with Crowe type III dislocated hips underwent 28 THAs using bone graft from the femoral neck between 1979 and 2000. A Charnley cemented acetabular component was placed at the level of the true acetabulum after preparation with bone grafting. All patients were female with a mean age at the time of surgery of 54 years (35 to 68). A total of 18 patients (21 THAs) were followed for a mean of 27.2 years (20 to 33) after the operation.Aims
Methods
With recent progress in cancer treatment, the number of advanced-age patients with spinal metastases has been increasing. It is important to clarify the influence of advanced age on outcomes following surgery for spinal metastases, especially with a focus on subjective health state values. We prospectively analyzed 101 patients with spinal metastases who underwent palliative surgery from 2013 to 2016. These patients were divided into two groups based on age (< 70 years and ≥ 70 years). The Eastern Cooperative Oncology Group (ECOG) performance status (PS), Barthel index (BI), and EuroQol-5 dimension (EQ-5D) score were assessed at study enrolment and at one, three, and six months after surgery. The survival times and complications were also collected.Aims
Methods
The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad).Aims
Methods
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods
To determine ten-year failure rates following 36 mm metal-on-metal
(MoM) Pinnacle total hip arthroplasty (THA), and identify predictors
of failure. We retrospectively assessed a single-centre cohort of 569 primary
36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012
according to Medicines and Healthcare Products Regulation Agency
recommendations. All-cause failure rates (all-cause revision, and
non-revised cross-sectional imaging failures) were calculated, with predictors
for failure identified using multivariable Cox regression.Aims
Patients and Methods
The interest in unicompartmental knee arthroplasty (UKA) for
medial osteoarthritis has increased rapidly but the long-term follow-up
of the Oxford UKAs has yet to be analysed in non-designer centres.
We have examined our ten- to 15-year clinical and radiological follow-up
data for the Oxford Phase III UKAs. Between January 1999 and January 2005 a total of 138 consecutive
Oxford Phase III arthroplasties were performed by a single surgeon
in 129 patients for medial compartment osteoarthritis (71 right
and 67 left knees, mean age 72.0 years (47 to 91), mean body mass
index 28.2 (20.7 to 52.2)). Both clinical data and radiographs were
prospectively recorded and obtained at intervals. Of the 129 patients,
32 patients (32 knees) died, ten patients (12 knees) were not able
to take part in the final clinical and radiological assessment due
to physical and mental conditions, but via telephone interview it
was confirmed that none of these ten patients (12 knees) had a revision
of the knee arthroplasty. One patient (two knees) was lost to follow-up.Aims
Patients and Methods
We investigated whether blood metal ion levels could effectively
identify patients with bilateral Birmingham Hip Resurfacing (BHR)
implants who have adverse reactions to metal debris (ARMD). Metal ion levels in whole blood were measured in 185 patients
with bilateral BHRs. Patients were divided into those with ARMD
who either had undergone a revision for ARMD or had ARMD on imaging
(n = 30), and those without ARMD (n = 155). Receiver operating characteristic
analysis was used to determine the optimal thresholds of blood metal
ion levels for identifying patients with ARMD.Aims
Patients and Methods