The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.Aims
Methods
This study evaluates the association between consultant and hospital volume and the risk of re-revision and 90-day mortality following first-time revision of primary hip arthroplasty for aseptic loosening. We conducted a cohort study of first-time, single-stage revision hip arthroplasties (RHAs) performed for aseptic loosening and recorded in the National Joint Registry (NJR) data for England, Wales, Northern Ireland, and the Isle of Man between 2003 and 2019. Patient identifiers were used to link records to national mortality data, and to NJR data to identify subsequent re-revision procedures. Multivariable Cox proportional hazard models with restricted cubic splines were used to define associations between volume and outcome.Aims
Methods
The aims of this study were to validate the outcome of total elbow arthroplasty (TEA) in patients with rheumatoid arthritis (RA), and to identify factors that affect the outcome. We searched PubMed, MEDLINE, Cochrane Reviews, and Embase from between January 2003 and March 2019. The primary aim was to determine the implant failure rate, the mode of failure, and risk factors predisposing to failure. A secondary aim was to identify the overall complication rate, associated risk factors, and clinical performance. A meta-regression analysis was completed to identify the association between each parameter with the outcome.Aims
Methods
Osseointegrated prosthetic limbs allow better mobility than socket-mounted prosthetics for lower limb amputees. Fractures, however, can occur in the residual limb, but they have rarely been reported. Approximately 2% to 3% of amputees with socket-mounted prostheses may fracture within five years. This is the first study which directly addresses the risks and management of periprosthetic osseointegration fractures in amputees. A retrospective review identified 518 osseointegration procedures which were undertaken in 458 patients between 2010 and 2018 for whom complete medical records were available. Potential risk factors including time since amputation, age at osseointegration, bone density, weight, uni/bilateral implantation and sex were evaluated with multiple logistic regression. The mechanism of injury, technique and implant that was used for fixation of the fracture, pre-osseointegration and post fracture mobility (assessed using the K-level) and the time that the prosthesis was worn for in hours/day were also assessed.Aims
Methods
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods