Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

TRIPLE TAPERED CEMENTED POLISHED STEM IN TOTAL HIP ARTHROPLASTY. A REVIEW OF 1008 CASES USING THE C-STEM WITH A MINIMUM OF 5 YEARS CLINICAL AND RADIOLOGICAL FOLLOW-UP.



Abstract

The Triple-tapered cemented polished C-Stem has evolved from the study of long-term results of the Charnley low-frictional torque arthroplasty when the first fractured stem and then proximal strain shielding of the femur and stem loosening were identified as the continuation of the same process- the lack or loss of proximal stem support.

The concept, design and the surgical technique cater for a limited slip of the C-stem within the cement mantle transferring the load more proximally. With a follow-up past 12 years and 4063 primary procedures there have been no revisions for aseptic stem loosening and no stem is radiologically loose.

We have reviewed 1008 primary C-Stem hip arthroplasties performed by 23 surgeons with a minimum of 5 years clinical and radiological follow-up. The mean follow-up was 7 years (range, 5 – 12) and the mean age at surgery was 57 years (range (15 – 85). In 58% the underlying pathology was primary osteoarthritis, 20% congenital dysplasia, 10% quadrantic head necrosis, 5% rheumatoid arthritis, 5% slipped upper femoral epiphysis and 4% protrusio acetabulae.

The concept of the triple tapered stem is validated radiologically with an improved proximal femoral bone stock in over 20% of cases and a maintained bone stock in 60%.

There were no post-operative complications within 1 year in 87% and no late complications (after 1 year) in 91%. The main late complications were 3.9% aseptic cup loosening, 1% infection and 0.8% dislocation. There were no aseptic loose stems.

Twenty-eight hips have been revised (2.8%), 3 for infection, 2 for dislocation and 23 for aseptic cup loosening. There were no revisions for aseptic stem loosening.

The results support the concept but place a demand on the understanding of the technique and its execution at surgery.

Correspondence should be addressed to Ms Larissa Welti, Scientific Secretary, EFORT Central Office, Technoparkstrasse 1, CH-8005 Zürich, Switzerland