Abstract
The Triple-tapered cemented polished C-Stem has evolved from the study of long-term results of the Charnley low-frictional torque arthroplasty when the first fractured stem and then proximal strain shielding of the femur and stem loosening were identified as the continuation of the same process- the lack or loss of proximal stem support.
The concept, design and the surgical technique cater for a limited slip of the C-stem within the cement mantle transferring the load more proximally. With a follow-up past 12 years and 4063 primary procedures there have been no revisions for aseptic stem loosening and no stem is radiologically loose.
We have reviewed 1008 primary C-Stem hip arthroplasties performed by 23 surgeons with a minimum of 5 years clinical and radiological follow-up. The mean follow-up was 7 years (range, 5 – 12) and the mean age at surgery was 57 years (range (15 – 85). In 58% the underlying pathology was primary osteoarthritis, 20% congenital dysplasia, 10% quadrantic head necrosis, 5% rheumatoid arthritis, 5% slipped upper femoral epiphysis and 4% protrusio acetabulae.
The concept of the triple tapered stem is validated radiologically with an improved proximal femoral bone stock in over 20% of cases and a maintained bone stock in 60%.
There were no post-operative complications within 1 year in 87% and no late complications (after 1 year) in 91%. The main late complications were 3.9% aseptic cup loosening, 1% infection and 0.8% dislocation. There were no aseptic loose stems.
Twenty-eight hips have been revised (2.8%), 3 for infection, 2 for dislocation and 23 for aseptic cup loosening. There were no revisions for aseptic stem loosening.
The results support the concept but place a demand on the understanding of the technique and its execution at surgery.