Cemented, polished, tapered stems have produced excellent results, but some early failures occur in younger patients. The CPS-Plus stem (Plus Orthopedics AG, Switzerland) is a polished double taper with rectangular cross section for improved rotational stability. A unique proximal stem centraliser increases cement pressurisation, assists alignment and creates an even cement mantle.
Radiostereometric analysis has demonstrated linear subsidence in a vertical plane, without any rotation or tilt. These features should improve implant durability. Midterm (5 years) results of a prospective international multicentre study are presented.
Materials and Methods: 222 patients (230 hips) were recruited to this IRB-approved study at three centres in the UK and two in Norway. Clinical and radiographic outcomes were assessed at regular intervals.
Results: 160 hips in 153 patients were available for full clinical and radiographic evaluation. 27 patients have died, 30 patients were unable to attend (outcome known) and 12 patients have not reached 5 years follow-up.
The mean Harris hip score improved from 42 preoperatively to 91. There have been no revisions for aseptic loosening and none of the stems have radiographic evidence of loosening. There has been one revision for deep sepsis. With revision for aseptic loosening as an endpoint, stem survivorship is 100%.
Conclusion: The design of the CPS-Plus stem attempts to address the issues of cement pressurization, rotational stability, and subsidence. Earlier laboratory studies have now been supplemented by this clinical evaluation, performed in a number of different centres by several surgeons, and the midterm results are very encouraging.