Full Access

CEMENTLESS METAL-ON-METAL HIP REPLACEMENT IN PATIENTS LESS THAN FIFTY YEARS OF AGE: COMPARISON WITH A MATCHED CONTROL GROUP USING CERAMIC-ON-POLYETHYLENE AFTER A MINIMUM 5-YEAR FOLLOW-UP.



Abstract

Thirty-nine cementless hip replacements using metal-on-metal articulation were consecutively implanted in thirty patients less than fifty years of age and compared with a matched control group of cementless replacements using ceramic-on-polyethylene articulation. The Harris hip score at follow-up (minimum five years) for the metal-on-metal was 94.9 (range, 74–100). After the same follow-up, the results of the ceramic-on-polyethylene were significantly worse: nine osteolyses and seven surgical revisions related to wear. Five-year survival rates were 97% +/− 2% for the ceramic-on-polyethylene and 100% for the metal-on-metal. The metal-on-metal may be recommended to prevent wear problems in younger and more active patients.

The aim of the current study was to assess the results of metal-on-metal articulating components inserted as a primary hip replacement in patients under the age of fifty, comparated with a matched control group using ceramic-on-polyethylene.

Patients and methods

The inclusion of patients was: under fifty years of age and a diagnosis of arthrosis or necrosis of the femoral head.

Femoral stem and cup migration was detected. A variation over five millimeters between the follow-up radiographs was considered as migration.

At the follow-up in 2003, the cobalt concentrations in the whole blood were assessed in the metal-on-metal cohort. The detection limit of cobalt in the whole blood was 0.06 μg/L.

None of the components had migration.

At a mean follow-up of sixty-nine months, the median concentration of cobalt in the whole blood was 0.62μg/L. Only eight patients had cobalt levels greater than 1 μg/L.

Considering a reoperation with the exchange of one of the components as end point, the five year survival rates were 100% for the metal-on-metal group and 97% + 2 for the ceramic-on-polyethylene group. Our study suggests that the metal-on-metal articulation gives a significant improvement in terms of resistance to wear when compared with these conventional bearing components. Our results suggest the metal-on-metal articulation with cementless components can be recommended in the young and active patient to prevent the occurrence of wear and osteolysis. A careful assessment of patients with high levels of whole-blood cobalt should be performed.

Funding: Aucun

Correspondence should be addressed to Cynthia Vezina, Communications Manager, COA, 4150-360 Ste. Catherine St. West, Westmount, QC H3Z 2Y5, Canada