OUTCOME OF COPELAND SURFACE REPLACEMENT SHOULDER ARTHROPLASTY



Abstract

The Copeland Shoulder Arthroplasty is a cementless, pegged humeral head surface replacement. The design is based on the principle of minimal bone resection and has been in clinical use since 1986. The only published series to date, that of Levy and Copeland, reported results for 103 patients which were comparable to those obtained with stemmed implants. We report the outcome at our institution using the same prosthesis with a hydroxyapatite coating.

81 shoulders (74 patients) underwent resurfacing hemiarthroplasty through an anterior deltopectoral approach. Preoperative diagnoses were: osteoarthritis (39), rheumatoid arthritis (29), rotator cuff arthropathy (1), post-traumatic arthrosis (2). They were followed for an average 28 months. 10 were lost to follow-up (8 deaths).

Constant scores improved from a mean preoperative figure of 15.7 to 54.0 (p< 0.01) at last follow-up. For rheumatoid arthritis and osteoarthritis the scores improved from 15.2 to 50.4 (p< 0.01) and 16.0 to 55.4 (p,< 0.01) respectively.

There was a 13% complication rate with one case requiring revision for loosening to a stemmed implant. Most were cases requiring subsequent acromioplasty. In one case the glenoid rim was fractured during head dislocation. There was a low rate of perioprosthetic radiolucency (4.2%) which may relate to the hydroxyappatite coating within the shell of the prosthesis. Ipsilateral stemmed elbow replacement was performed in some cases without a double stress riser effect. Periprosthetic humeral neck fracture as managed non-operatively with uneventful union.

Conclusion: The good outcome reported in Copeland’s own series has been replicated in the early to medium term at our institution. The surface replacement system is simpler, accurate and preserves bone stock.

The abstracts were prepared by Mr Tim Briggs. (Editoral Secretary 2003/4) Correspondence should be addressed to him at Lane Farm, Chapel Lane, Totternhoe, Dunstable, Bedfordshire LU6 2BZ, United Kingdom