POSITIONAL MRI CHANGES IN THE LUMBAR SPINE FOLLOWING INSERTION OF A NOVEL INTERSPINOUS PROSCESS DISTRACTION DEVICE



Abstract

Purpose: To measure the effect of the X-Stop interspinous distraction device on spinal canal, exit foramina, and disc height dimensions at the operated level; and adjacent segment endplate angle, and lumbar spine movement in patients with symptomatic lumbar spinal stenosis using upright MRI.

Methods /Results: 14 patients (9 M;5 F) were scanned before and six months after operation. Age ranged from 57 to 88 years. All had symptomatic lumbar spinal stenosis- single level- 9 (L2/3-1; L3/4-1; L4/5-7); double level 5 (L3/4, L4/5).

Images were taken in sitting flexed, extended, neutral, and standing. The total range of motion of the lumbar spine and of the individual segments were measured, along with changes in disc height, areas of the exit foramina, and dural sac.

The mean area of the dural sac at the operated levels increased from 62.46mm2 to 77.69mm2 (p=0.004) in the standing posture and from 70.85mm2 to 94.62mm2 (p=0.019) in extension postoperatively. The area of the exit foramina in extension increased from 83.57mm2 to 107.88mm2 (p=0.002) on the left side and from 83.77mm2 to 108.69mm2 (p=0.012) on the right. The overall changes in the range of movement of the individual segments or of the lumbar spine were statistically insignificant.

Conclusions: This is the first study carried out using an upright MRI scanner in patients with lumbar spinal stenosis. The X-Stop device increases the cross-sectional area of the spinal canal and exit foramina by distracting the spinous processes of the operated level without significantly affecting overall posture of the lumbar spine.