PREDICTION OF CHARNLEY FEMORAL STEM ASEPTIC LOOSENING BY EARLY POST-OPERATIVE RADIOLOGICAL FEATURES

7th Congress of the European Federation of National Associations of Orthopaedics and Traumatology, Lisbon - 4-7 June, 2005



Abstract

We describe the association between immediate postoperative radiological appearances and early aseptic failure of THA having compensated for the methodological flaws in previous similar studies. 63 hips were entered into the aseptic failure group and 138 into the control group. Alignment of the femoral stem was not associated with failure (p=0.283). Thickness of the cement mantle was associated with failure in Gruen zones 6 (p=0.040) and 7 (p=0.003). A significant association for the presence of radiolucent lines was found for Gruen zones 3 (p=0.0001) and 5 (p=0.0001). Grade of cementation was associated with failure for Barrack grades C (p=0.001) and D (p=0.001). This study has demonstrated that easily applied radiological criteria can be used to identify at risk THAs from the immediate post-operative AP radiograph.