Abstract
Background and purposes of study The objective of this study is to report the preliminary clinical data from a stand-alone interbody fusion cage (Stabilis™) from two centres.
Methods and results This was a retrospective group comparative study. Anterior lumbar interbody fusion was performed using Brantigan (n=6) or Stabilis™ cages (n=19). Clinical assessment was performed using standardised questionnaires. Lumbar flexion/extension views were used to assess union and implant subsidence. There were no significant differences between the groups for ODI, VAZ, MZ and MSPQ pre-operatively. The Stabilis group showed an improvement in ODI (pre: 49.4; post: 39.3; p=0.024), VAS (pre: 76.1; post: 47.8; p< 0.01) and MZ (pre 32.9; post: 20.6; p< 0.01).
There was no statistical improvement in the Brantigan cage group. Despite clinical improvement, five patients in the Stabilis group failed to unite and six demonstrated subsidence of the implant. The relationship between non-union and subsidence was statistically significant (p = 0.017). Furthermore, the change in ODI between patients who united and those who did not was both statistically significant (p=0.03) and the difference in mean ODI between the two groups was considerable (21%).
Conclusions: Stand-alone cages show promise, involve a shorter operating time and less tissue trauma for the patient. This study has identified a high rate of non-union and implant subsidence. This did not appear to be clinically relevant.
Correspondence should be addressed to SBPR c/o Royal college of Surgeons, 35 - 43 Lincoln’s Inn Fields, London WC2A 3PN