Abstract
Introduction: This study investigates the influence of surgical approach on the early migration of a cemented Total Hip Replacement (THR), assessed by Roentgen Stereo-photogrammetric Analysis (RSA). Rapid early migration has been correlated with premature implant failure.
Method: 46 patients awaiting THR were allocated into Posterior Approach (PA) and Lateral Approach (LA) groups. RSA was used to assess post-operative stem position and migration at 1 year. Post-operative gait was examined for a subgroup of patients.
Results: There was no difference in initial stem position within the medulla. Significant differences in distal migration were found; (LA 0.78±0.07mm and PA 1.27±0.15mm), p< 0.01. Posterior head migration for the PA group (0.78±0.18mm) was twice that for the LA group (0.46±0.08mm), p< 0.05. The PA group showed greater medial migration of the prosthesis tip (0.62±0.24mm), compared to the LA group (0.11±0.08mm), p< 0.05. During gait, the legs of the PA group were relatively internally rotated when compared to those of the LA group.
Discussion: The PA implants rotate internally and into valgus further than the LA implants. The starting positions of the prostheses were the same with similar cement and prostheses characteristics. We can infer that surgical approach and therefore muscle function have a significant effect upon component migration. During gait, the PA group has relatively internally rotated legs, probably due to short external rotator weakness. Hence, the internally rotating torque on the femoral stem will be larger, explaining the rapid migration in this direction. With the LA approach, the abductors are likely to be damaged. Therefore, the PA joint reaction force will be more laterally directed, explaining the higher coronal plane rotation rates seen in this group.
Conclusion: Profound differences exist in early migration, between the LA and PA. Different muscle function may account for this and probably influences long-term outcome.
The abstracts were prepared by Nico Verdonschot. Correspondence should be addressed to him at Orthopaedic Research Laboratory, University Medical Centre, PO Box 9101, 6500 HB Nijmegen, The Netherlands.