Abstract
This early study examines the influence of a wider shoulder on the 1 year migration of a cemented, polished, tapered stem, using RSA.
Polished, tapered stems (PTS) have excellent 10 year survival rates. RSA studies have demonstrated that these devices subside about 1 mm / year. Small amounts of subsidence are beneficial in stabilising a stem. Stem rotation (measured as posterior head migration) within the cement mantle is probably a more important mechanism of failure than subsidence. Stems with a wider proximal portion are thought to better resist rotation. The CPS (Endoplus, UK) is such a device; here we compare its’ stability with that of the Exeter.
20 patients received the CPS-plus stem and underwent RSA examinations at 3, 6 and 12 months postoperatively. The Exeter 1 year migration data was used as a comparison. Both groups underwent a Hardinge approach with CMW3G cement.
Both stems subsided about 1mm. The CPS showed less medio-lateral and A-P movement of the proximal stem than the Exeter over 1 year, as shown below:
The CPS internally rotates less than the Exeter, as demonstrated by the smaller amount of posterior head migration. It has a lateral flare of the shoulder; making its cross-section wider than the Exeter’s, this probably accounts for its’ greater resistance to rotation. The CPS also undergoes less medio-lateral proximal stem migration. Its’ lateral shoulder flare is probably responsible for this axial subsidence, as it prevents the shoulder from moving laterally whilst subsiding over the calcar.
A PTS with a broad proximal section is more stable, this may confer an increased survival advantage.
These abstracts were prepared by Mr Peter Kay. Correspondence should be addressed to him at The Hip Centre, Wrightington Hospital, Appley Bridge, Wigan, Lancashire WN6 9EP.