Abstract
The newer techniques of cementing aim to improve interlock between cement and bone around a femoral stem by combining high pressure and reduced viscosity. This may produce increased embolisation of fat and marrow leading to hypotension, impaired pulmonary gas exchange and death. For this reason the use of high pressures has been questioned.
We compared finger-packing with the use of a cement gun by measuring intramedullary pressures during the cementing of 31 total hip replacements and measuring physiological changes in 19 patients. We also measured pressure in more detail in a laboratory model.
In the clinical series the higher pressures were produced by using a gun, but this caused less physiological disturbance than finger-packing. The laboratory studies showed more consistent results with the gun technique, but for both methods of cementing the highest pressures were generated during the insertion of the stem of the prosthesis.