Advertisement for orthosearch.org.uk
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Laboratory comparison of the cannulated Herbert bone screw with ASIF cancellous lag screws



Download PDF

Abstract

The compression produced by and the resistance to pullout of the 6.5 mm cannulated Herbert screw were compared with those of ASIF headed screws. The latter were tested with and without washers and in the following sizes: 4.5 mm cortical, 6.5 mm cancellous with a 16 mm threaded segment, and 6.5 mm cancellous with a 32 mm threaded segment. Polyurethane foam was used as a substitute for cancellous bone and ASIF artificial bone for corticocancellous bone. The compression produced by a cancellous lag screw with a washer was significantly greater than that produced by a Herbert screw of equivalent size (p < 0.05). When the screws were tested using the corticocancellous composite the ASIF cancellous screw without a washer produced significantly greater compression (p < 0.05); when used with a washer the difference was highly significant (p < 0.001). The dual pitch Herbert screw is not appropriate for the management of fractures in which compression is of greater importance than the need to avoid prominence of the screw head.

For access options please click here